基于定位像的骨密度定量:能量积分探测器CT与光子计数探测器CT的准确性与精密度

Bone Mineral Density Quantification from Localizer Radiographs: Accuracy and Precision of Energy-integrating Detector CT and Photon-counting Detector CT.
PubMed 文献 · Radiology · 2020-11-03

研究用校准体模和猪腰椎体模,比较能量积分探测器CT(双定位像、不同管电压)与光子计数探测器CT(单定位像、不同能量阈值)测量骨密度的准确性与精密度,并以双能X线吸收测量仪(DXA)为参考标准。两种CT系统测量误差均小于DXA,精密度高,动物体模中两种CT系统间无显著差异(P = .65),提示定位像CT骨密度测量有望用于机会性骨质疏松筛查。

要点

  • 能量积分探测器CT用两张不同管电压定位像、光子计数探测器CT用单张不同能量阈值定位像即可定量骨密度。
  • 校准体模中光子计数探测器CT的骨密度测量误差范围为-1.6%至1.6%,能量积分探测器CT为-2.3%至-1.7%,DXA为-5%至-1.8%。
  • 动物体模中两种CT系统骨密度测量误差均很小,且两者间无显著差异(P = .65)。

核心结论

  • 实验证据提示,能量积分探测器CT采用不同管电压的两张定位像、光子计数探测器CT采用不同能量阈值的单张定位像测量骨密度均准确且精密度高。

研究设计与样本

  • 实验研究,使用校准体模和含腰椎的猪体模,分别以DXA扫描仪、临床能量积分探测器CT扫描仪和原型光子计数探测器CT扫描仪成像,并以DXA为参考标准,采用非参数检验。

诊断标准与关键数值

  • 校准体模中骨密度测量误差:DXA为-5%至-1.8%,能量积分探测器CT为-2.3%至-1.7%,光子计数探测器CT为-1.6%至1.6%,两种CT系统误差均小于DXA(均P < .05)。

临床应用与阅片要点

  • 作者认为基于定位像的骨密度测量可能有助于机会性骨质疏松筛查;阅片时可关注定位像上骨密度定量结果,但需注意其准确性依赖管电压或能量阈值组合。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background Bone mineral density (BMD) could be derived from CT localizer radiographs and could potentially enable opportunistic osteoporosis screening. Purpose To assess the accuracy and precision of BMD measurement using two localizer radiographs obtained with energy-integrating detector CT and a single localizer radiograph obtained with photon-counting detector CT. Materials and Methods A calibration phantom and a porcine phantom with lumbar vertebrae were imaged with a dual-energy x-ray absorptiometry (DXA) scanner, a clinical energy-integrating detector CT scanner, and a prototype photon-counting detector CT scanner. Two localizer radiographs at different combinations of tube voltages were obtained with energy-integrating detector CT, and one localizer radiograph was obtained with photon-counting detector CT using different energy thresholds. BMD was calculated for all three approaches and compared with the known specifications in the calibration phantom. In the animal phantom, BMDs from both CT systems were compared with those from the DXA scanner (the reference standard). Accuracy was defined as the measurement error of BMD (ΔBMD), and precision was defined as the coefficient of variation (in percentage). Radiation doses were estimated. Nonparametric tests were applied. Results In the calibration phantom, ΔBMD was smaller with both CT systems compared with the DXA scanner (both P < .05). ΔBMD ranged from -5% to -1.8% for DXA, from -2.3% to -1.7% for energy-integrating detector CT, and from -1.6% to 1.6% for photon-counting detector CT. Precision (range, 0.3%-2.8%) was high for both CT systems. In the animal phantom, ΔBMD ranged from -0.6% to 0.1% for energy-integrating detector CT and from -0.1% to 0.6% for photon-counting detector CT, with no significant differences between CT systems (P = .65). The dose-area product in the animal phantom was 4.6 cGy ∙ cm2 for DXA, 3.5-11.5 cGy ∙ cm2 for energy-integrating detector CT, and 7.2-11.2 cGy ∙ cm2 for photon-counting detector CT, depending on tube voltage and energy threshold combination. Conclusion Experimental evidence suggests that bone mineral density measurements are accurate and precise using two localizer radiographs at different tube voltages from energy-integrating detector CT and a single localizer radiograph with different energy thresholds from photon-counting detector CT. © RSNA, 2020 Online supplemental material is available for this article. See also the editorial by Pourmorteza in this issue.