文献速递外周血管介入DSA

基于设备剂量学信息和峰值皮肤剂量优化的血管透视引导介入放射学预警阈值

Optimized radiological alert thresholds based on device dosimetric information and peak skin dose in vascular fluoroscopically guided intervention.
PubMed 文献 · Eur Radiol · 2020-11-06

本研究纳入108例血管透视引导介入中至少达到一项NCRP阈值的患者,以峰值皮肤剂量(PSD)为参考,评估空气比释动能、剂量面积乘积(KAP)和透视时间与PSD的相关性,并提出优化阈值。主成分分析显示仅KAP和空气比释动能与PSD相关,透视时间不适宜预测PSD>3 Gy;优化阈值分别为KAP 85,451 μGy·cm²、空气比释动能2938 mGy和透视时间41 min。

要点

  • 透视时间与峰值皮肤剂量不相关,不宜单独用于预测PSD>3 Gy。
  • KAP和空气比释动能与PSD相关,优化阈值为85,451 μGy·cm²和2938 mGy。
  • 优化阈值预测PSD>3 Gy的敏感度/特异度:KAP 67.6/93.0,空气比释动能97.3/81.7。

诊断标准与关键数值

  • NCRP报告第168号建议在透视引导介入中达到以下任一阈值时监测患者:空气比释动能>5 Gy、KAP>500 Gy·cm²、透视时间>60 min或PSD>3 Gy。
  • 优化阈值:KAP 85,451 μGy·cm²、空气比释动能2938 mGy、透视时间41 min。
  • 优化阈值预测PSD>3 Gy的敏感度/特异度:KAP 67.6/93.0,空气比释动能97.3/81.7,透视时间62.2/62.0。
  • NCRP阈值预测PSD>3 Gy的敏感度/特异度:KAP 97.3/16.9,空气比释动能40.5/95.4,透视时间21.6/74.7。

临床应用与阅片要点

  • 在血管透视引导介入中,可依据KAP和空气比释动能的优化阈值(85,451 μGy·cm²和2938 mGy)触发患者随访。
  • 透视时间不宜作为预测PSD>3 Gy的指标。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The National Council on Radiation Protection (NCRP) report no. 168 recommended that during fluoroscopically guided interventions (FGIs), each patient should be monitored when one of the following thresholds is reached: an air kerma > 5 Gy, a kerma area product (KAP) > 500 Gy.cm2, a fluoroscopy time > 60 min, or a peak skin dose (PSD) > 3 Gy. Whereas PSD is the most accurate metric regarding the prevention of radiological risks, it remains the most difficult parameter to assess. We aimed to evaluate the relevance of the other, more accessible metrics and propose new optimized threshold (OT) for improved patient follow-up.

METHODS: Overall, 108 patients who underwent FGI in which at least one NCRP threshold was reached and PSD was measured were considered. The correlation between all metrics was assessed using principal component analysis (PCA). ROC curves and the sensitivity/specificity of both NCRP and OT to predict PSD > 3 Gy were evaluated.

RESULTS: The PCA shows that FGI can be decomposed with two components based on time and dose variables. Only KAP and kerma were correlated with PSD. The overall sensitivity and specificity of the new OT regarding KAP (67.6/93.0), kerma (97.3/81.7), and time (62.2/62.0) were better compared with NCRP thresholds (97.3/16.9, 40.5/95.4, and 21.6/74.7).

CONCLUSIONS: This study shows that fluoroscopy time is not a relevant metric when used to predict PSDs > 3 Gy. By adapting KAP and kerma thresholds to predict PSD over 3 Gy, patient follow-ups following vascular FGI can be improved.

KEY POINTS: • In vascular fluoroscopically guided interventions, principal component analysis demonstrates that between fluoroscopy time, KAP, and kerma, only the two last were correlated to the peak skin dose. • Optimized thresholds replacing NRCP ones obtained with ROC curves analysis were 85,451 μGy.cm2, 2938 mGy, and 41 min for KAP, kerma, and fluoroscopy time respectively. • Improvements to trigger patient follow-up after vascular fluoroscopically guided interventions may be obtained by using the optimized thresholds.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Sas N, Magnier F, Pouget E, et al. Optimized radiological alert thresholds based on device dosimetric information and peak skin dose in vascular fluoroscopically guided intervention. Eur Radiol. 2020. PMID: 33156387. doi:10.1007/s00330-020-07422-3