实验性体外模型中常规CT上液体栓塞剂成像伪影的评估
本研究在体外实验模型中,将Onyx 18、Squid全系列、PHIL 25%、PHIL LV及n-BCA与碘化油混合物填充于二甲基亚砜兼容管并置入CT成像体模,定量与定性评估常规CT伪影。Onyx 18伪影最多,n-BCA、PHIL 25%和PHIL LV最少;Squid伪影较PHIL多,以碘为不透射线成分的液体栓塞剂伪影少于乙烯-乙烯醇基液体栓塞剂。
要点
- Onyx 18在常规CT上产生的伪影最多,n-BCA、PHIL 25%和PHIL LV产生的伪影最少。
- Squid在低粘度和超低粘度版本中产生的伪影均多于PHIL(如Squid 18均值30.3 HU vs PHIL 25%均值10.6 HU,P < .001)。
- 以碘为不透射线成分的液体栓塞剂相比乙烯-乙烯醇基液体栓塞剂产生更少CT伪影。
核心结论
- 在该系统体外分析中,最常用的液体栓塞剂在常规CT上产生的成像伪影存在相关差异。
- 乙烯-乙烯醇基液体栓塞剂相比以碘为不透射线成分的液体栓塞剂产生更多伪影。
研究设计与样本
- 实验性体外模型研究;二甲基亚砜兼容管分别填充Onyx 18、Squid全系列、PHIL 25%、PHIL LV及n-BCA与碘化油混合物,每种栓塞剂n = 10。
- 将管插入CT成像体模后采集常规CT图像;伪影定量采用环形ROI内CT值(HU)并借助为本研究专门设计的定制软件分析,定性采用5分制评分。
诊断标准与关键数值
- 定量分析:Squid 18均值±标准差30.3 ± 9.7 HU vs PHIL 25%均值±标准差10.6 ± 0.8 HU,P < .001。
- Squid与PHIL在低粘度和超低粘度版本中,Squid产生的伪影均多于PHIL。
- Squid标准与低密度变体之间的差异仅在Squid 12的定量分析中观察到。
- Squid与PHIL不同浓度之间无统计学差异。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Endovascular embolization using liquid embolic agents is a safe and effective treatment option for AVMs and dural arteriovenous fistulas. The aim of this study was to assess the degree of artifact inducement by the most frequently used liquid embolic agents in conventional CT in an experimental in vitro model.
MATERIALS AND METHODS: Dimethyl-sulfoxide-compatible tubes were filled with the following liquid embolic agents (n = 10, respectively): Onyx 18, all variants of Squid, PHIL 25%, PHIL LV, and n-BCA mixed with iodized oil. After inserting the tubes into a CT imaging phantom, we acquired images. Artifacts were graded quantitatively by the use of Hounsfield units in a donut-shaped ROI using a customized software application that was specifically designed for this study and were graded qualitatively using a 5-point scale.
RESULTS: Quantitative and qualitative analyses revealed the most artifacts for Onyx 18 and the least artifacts for n-BCA, PHIL 25%, and PHIL LV. Squid caused more artifacts compared with PHIL, both for the low-viscosity and for the extra-low-viscosity versions (eg, quantitative analysis, Squid 18: mean ± SD, 30.3 ± 9.7 HU versus PHIL 25%: mean ± SD, 10.6 ± 0.8 HU; P < .001). Differences between the standard and low-density variants of Squid were observed only quantitatively for Squid 12. There were no statistical differences between the different concentrations of Squid and PHIL.
CONCLUSIONS: In this systematic in vitro analysis investigating the most commonly used liquid embolic agents, relevant differences in CT imaging artifacts could be demonstrated. Ethylene-vinyl alcohol-based liquid embolic agents induced more artifacts compared with liquid embolic agents that use iodine as a radiopaque component.