婴儿髋关节发育不良度量指标的3-D超声成像可重复性研究
本研究提出并评估了自动计算的3-D超声衍生发育不良度量指标,基于40例婴儿髋关节检查的临床数据验证,结果显示两个关键3-D指标较2-D对应指标的变异性降低约70%。
要点
- 自动计算的3-D超声发育不良度量指标较2-D指标变异性降低约70%。
- 方法基于随机森林学习、骨分割和3-D重建策略自动提取度量指标。
- 在40例婴儿髋关节检查的真实临床数据上验证了所提方法。
临床应用与阅片要点
- 3-D超声自动衍生发育不良度量指标可减少检查间变异性,有助于提高诊断可重复性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Developmental dysplasia of the hip is a hip abnormality that ranges from mild acetabular dysplasia to irreducible femoral head dislocations. While 2-D B-mode ultrasound (US)-based dysplasia metrics or disease metrics are currently used clinically to diagnose developmental dysplasia of the hip, such estimates suffer from high inter-exam variability. In this work, we propose and evaluate 3-D US-derived dysplasia metrics that are automatically computed and demonstrate that these automatically derived dysplasia metrics are considerably more reproducible. The key features of our automatic method are (i) a random forest-based learning technique to remove regions across the coronal axis that do not contain bone structures necessary for dysplasia-metric extraction, thereby reducing outliers; (ii) a bone segmentation method that uses rotation-invariant and intensity-invariant filters, thus remaining robust to signal dropout and varying bone morphology; (iii) a novel slice-based learning and 3-D reconstruction strategy to estimate a probability map of the hypoechoic femoral head in the US volume; and (iv) formulae for calculating the 3-D US-derived dysplasia metrics. We validate our proposed method on real clinical data acquired from 40 infant hip examinations. Results show a considerable (around 70%) reduction in variability in two key 3-D US-derived dysplasia metrics compared with their 2-D counterparts.
引用
- Quader N, Hodgson AJ, Mulpuri K, et al. 3-D Ultrasound Imaging Reliability of Measuring Dysplasia Metrics in Infants. Ultrasound Med Biol. 2020. PMID: 33239155. doi:10.1016/j.ultrasmedbio.2020.08.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。