二维数字立体定向与数字乳腺断层摄影引导乳腺活检的辐射剂量比较
该回顾性研究比较了数字乳腺断层摄影(DBT)引导与数字乳腺摄影(DM)引导乳腺活检的采集图像数量和辐射剂量。结果显示,DBT引导活检的平均采集图像数和平均腺体剂量均低于DM引导,且对于仅钙化病灶,DBT引导的平均腺体剂量也显著低于DM引导(10.7 mGy vs 22.3 mGy,p < 0.001)。
要点
- DBT引导乳腺活检的平均采集图像数为4.3幅,低于DM引导的9.3幅(双单元)和10.9幅(单DM单元)。
- DBT引导活检的平均手术平均腺体剂量为10.18 mGy,低于DM引导的22.06 mGy(双单元)和28.77 mGy(单DM单元)。
- 对于仅钙化病灶,DBT引导的平均腺体剂量为10.7 mGy,显著低于DM引导的22.3 mGy(p < 0.001)。
诊断标准与关键数值
- DBT引导活检的平均采集图像数为4.3幅,DM引导(双单元)为9.3幅,DM引导(单DM单元)为10.9幅。
- DBT引导活检的平均手术AGD为10.18 mGy,DM引导(双单元)为22.06 mGy,DM引导(单DM单元)为28.77 mGy。
- 仅钙化病灶:DBT引导的平均手术AGD为10.7 mGy,DM引导(双单元)为22.3 mGy(p < 0.001);DBT引导平均每例采集4.2幅图像,DM引导(双单元)为8.1幅。
临床应用与阅片要点
- 原文结论指出,DBT引导活检的图像采集数少于DM引导,因此DBT引导操作的整体剂量更低。
- 原文结论指出,DBT引导的剂量降低可能适用于所有病灶类型,甚至仅钙化病灶。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVE: Use of digital breast tomosynthesis (DBT) in breast imaging has necessitated DBT-guided biopsy, however, a single DBT acquisition may result in a greater radiation dose than a single DM acquisition. Our objective was to compare the number of images acquired and the resulting radiation dose of DBT versus DM-guided breast biopsies.
METHOD: All biopsies performed on our DM unit from 8/2016 to 1/2017 and on our DM-DBT unit from 8/2017 to 1/2018 were retrospectively reviewed. The number of image acquisitions, average glandular dose (AGD) per acquisition and per procedure were computed and stratified by guidance modality and lesion type.
RESULTS: 25 DM-guided biopsies were performed on the DM-only unit, 58 biopsies were performed with DM guidance on the dual unit (DM-DU) and 29 were performed with DBT. The average number of images acquisitions was 10.9 for DM-only unit biopsies, 9.3 images for DM-DU biopsies and 4.3 images for DBT-guided biopsies. Mean procedure AGD for DM-only unit biopsies was 28.77 mGy, versus 22.06 mGy for DM-DU and 10.18 mGy for DBT biopsies. Mean procedure AGD for biopsied calcification-only lesions was 22.3 mGy for DM-DU versus 10.7 mGy for DBT guidance (p < 0.001), with an average of 8.1 images per procedure for DM-DU versus 4.2 for DBT.
CONCLUSION: Fewer image acquisitions were obtained with DBT compared with DM guidance, therefore, the overall dose of DBT-guided procedures was less. The dose reduction obtained with DBT is possible across all lesion types, even for calcification-only lesions.
引用
- Amir T, Zuckerman SP, Barufaldi B, et al. Comparison of radiation dose between 2D digital stereotactic versus digital breast tomosynthesis-guided breast biopsies. Eur J Radiol. 2020. PMID: 33248401. doi:10.1016/j.ejrad.2020.109407
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。