同步多层技术加速读出分段回波平面成像用于乳腺病变诊断
这项前瞻性研究纳入96例患者共102个经组织病理学证实的乳腺病变(80个恶性、22个良性),在3T扫描仪上比较常规rs-EPI与SMS rs-EPI。SMS rs-EPI采集时间从4:27 min缩短至2:17 min,各参数相关性良好(ADC r=0.975),诊断性能与图像质量相当,SNR与CNR无差异。
要点
- SMS rs-EPI将采集时间从4:27 min缩短至2:17 min,图像质量与常规rs-EPI相当。
- SMS rs-EPI与常规rs-EPI的MK、MD、ADC值高度相关(r=0.908–0.975,p<0.01)。
- SMS rs-EPI鉴别乳腺良恶性病变的诊断性能与常规rs-EPI相当。
关键影像表现/征象
- SMS rs-EPI的采集时间为2:17 min,常规rs-EPI为4:27 min。
- 常规rs-EPI与SMS rs-EPI各参数呈极好线性关系:MK r=0.908、MD r=0.938、ADC r=0.975,p均<0.01。
- SMS rs-EPI的视觉图像质量与常规rs-EPI相当,读者间可靠性极好(ICC=0.851–0.940)。
- 常规rs-EPI与SMS rs-EPI的SNR和CNR均无差异(p>0.05)。
诊断标准与关键数值
- SMS rs-EPI与常规rs-EPI诊断性能相似,用于鉴别乳腺良恶性病变。
- 定量参数:MK、MD、ADC值在两种序列上均逐病变计算。
- 图像质量评估包括图像锐利度、几何畸变、病变清晰度、解剖结构、整体图像质量、SNR和CNR。
临床应用与阅片要点
- SMS rs-EPI可在保持与常规rs-EPI相当图像质量和诊断性能的前提下显著缩短乳腺MRI采集时间。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the feasibility and effectiveness of SMS rs-EPI for evaluating breast lesions.
METHODS: This prospective study was approved by IRB. Ninety-six patients had 102 histopathologically verified lesions (80 malignant and 22 benign) that were evaluated. Conventional rs-EPI and SMS rs-EPI data were acquired on a 3T scanner. Mean kurtosis (MK), mean diffusion (MD), and apparent diffusion coefficient (ADC) values were quantitatively calculated for each lesion on both sequences. Images were qualitatively and quantitatively analyzed with respect to image sharpness, geometric distortion, lesion conspicuity, anatomic structure, overall image quality, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). Student's t test, Pearson correlation, receiver operating characteristic curve, Wilcoxon rank sum test, and paired-sample t tests were used for statistical analysis.
RESULTS: Compared to conventional rs-EPI, the acquisition time of SMS rs-EPI was markedly reduced (2:17 min vs 4:27 min). Pearson's correlations showed excellent linear relationships for each parameter between conventional rs-EPI and SMS rs-EPI (MK, r = 0.908; MD, r = 0.938; and ADC, r = 0.975; p < 0.01 for all). Furthermore, SMS rs-EPI had similar diagnostic performance compared with conventional rs-EPI. SMS rs-EPI had comparable visual image quality as conventional rs-EPI, with excellent inter-reader reliability (ICC = 0.851-0.940). No differences existed between conventional rs-EPI and SMS rs-EPI for either SNR or CNR (p > 0.05).
CONCLUSIONS: Applying the SMS technique can significantly reduce the acquisition time and produce similar diagnostic accuracy while generating comparable image quality as the conventional rs-EPI.
KEY POINTS: • SMS rs-EPI reduces scan time from 4:27 min to 2:17 min compared with conventional rs-EPI. • SMS rs-EPI has a comparable diagnostic performance to conventional rs-EPI in the differentiation between malignant and benign breast lesions. • SMS rs-EPI demonstrates comparable image quality to conventional rs-EPI with shorter scan time.