文献速递腹部泌尿生殖技术与安全CT

低剂量双能CT用于结石成分表征:两代分滤板单源与双源双能CT的系统比较

Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT.
PubMed 文献 · Abdom Radiol (NY) · 2020-11-07

体外体模研究将32枚泌尿系结石嵌入明胶后置入水模,比较四种双能CT方法在3 mGy与6 mGy剂量下区分尿酸与非尿酸结石的噪声纹理和准确度。低剂量下A、B、C、D准确度分别为96.9%、96.9%、93.8%、93.8%,常规剂量下分别为100%、96.9%、96.9%、100%。除低剂量D外,厂商指定双能比阈值均为最优。

要点

  • 四种双能CT方法区分尿酸与非尿酸结石的准确度均较高,低剂量为93.8%–96.9%,常规剂量为96.9%–100%。
  • 第一代分滤板单源双能CT的噪声纹理更细,平均空间频率fav=0.27,其余三种方法为0.18–0.19。
  • 仅低剂量下第二代分滤板单源双能CT串行采集(D)使用优化阈值可将准确度从93.8%提升至100%。

关键影像表现/征象

  • 归一化噪声功率谱(nNPS)显示A、C、D的噪声纹理相近,平均空间频率fav范围为0.18–0.19。
  • B的噪声纹理更细,fav=0.27。

诊断标准与关键数值

  • 低剂量下区分尿酸与非尿酸结石的准确度:A为96.9%,B为96.9%,C为93.8%,D为93.8%。
  • 常规剂量下区分尿酸与非尿酸结石的准确度:A为100%,B为96.9%,C为96.9%,D为100%。
  • 除低剂量D外,厂商指定的双能比阈值均为最优;低剂量D使用优化阈值后准确度从93.8%提升至100%。

临床应用与阅片要点

  • 四种双能CT方法在低剂量和常规剂量下区分尿酸与非尿酸结石的准确度均高。
  • 第一代分滤板单源双能CT的噪声纹理更细。
  • 低剂量下第二代分滤板单源双能CT串行采集需使用优化阈值以保持准确度。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare noise texture and accuracy to differentiate uric acid from non-uric acid urinary stones among four different single-source and dual-source DECT approaches in an ex vivo phantom study.

METHODS: Thirty-two urinary stones embedded in gelatin were mounted on a Styrofoam disk and placed into a water-filled phantom. The phantom was imaged using four different DECT approaches: (A) dual-source DECT (DS-DE); (B) 1st generation split-filter single-source DECT (SF1-TB); (C) 2nd generation split-filter single-source DECT (SF2-TB) and (D) 2nd generation split-filter single-source DECT using serial acquisitions (SF2-TS). Two different radiation doses (3 mGy and 6 mGy) were used. Noise texture was compared by assessing the average spatial frequency (fav) of the normalized noise power spectrum (nNPS). ROC curves for stone classification were computed and the accuracy for different dual-energy ratio cutoffs was derived.

RESULTS: NNPS demonstrated comparable noise texture among A, C, and D (fav-range 0.18-0.19) but finer noise texture for B (fav = 0.27). Stone classification showed an accuracy of 96.9%, 96.9%, 93.8%, 93.8% for A, B, C, D for low-dose, respectively, and 100%, 96.9%, 96.9%, 100% for routine dose. The vendor-specified cutoff for the dual-energy ratio was optimal except for the low-dose scan in D for which the accuracy was improved from 93.8 to 100% using an optimized cutoff.

CONCLUSION: Accuracy to differentiate uric acid from non-uric acid stones was high among four single-source and dual-source DECT approaches for low- and routine dose DECT scans. Noise texture differed only slightly for the first-generation split-filter approach.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Nakhostin D, Sartoretti T, Eberhard M, et al. Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT. Abdom Radiol (NY). 2020. PMID: 33159558. doi:10.1007/s00261-020-02852-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。