主动脉CT血管造影:优化自动管电压选择设置以降低辐射剂量或对比剂用量的前瞻性随机试验

Computed Tomography Angiography of the Aorta-Optimization of Automatic Tube Voltage Selection Settings to Reduce Radiation Dose or Contrast Medium in a Prospective Randomized Trial.
PubMed 文献 · Invest Radiol · 2021-01-01

该单中心前瞻性随机对照试验将126例胸腹主动脉CTA患者分为三组:A组为120 kVp参考方案,B组为90 kVp降低辐射剂量并标准对比剂用量,C组为90 kVp标准辐射剂量并降低对比剂用量。结果显示B组尺寸特异性剂量估计值较A组降低34.3%(P<0.0001),C组对比剂用量较A组降低20.2%(P<0.0001),B组和C组的对比噪声比非劣效于A组,三组主观图像质量无显著差异(P=0.4)。

要点

  • 低kVp联合自动管电压选择滑块水平调整可在保证图像质量非劣效的前提下降低辐射剂量或对比剂用量。
  • B组(90 kVp、滑块水平11)较A组辐射剂量降低34.3%,对比噪声比非劣效。
  • C组(90 kVp、滑块水平3)较A组对比剂用量降低20.2%,对比噪声比非劣效。

核心结论

  • 自动管电压选择系统的滑块设置可调整以优化辐射剂量或对比剂用量,在低kVp主动脉CTA中实现非劣效的图像质量。

研究设计与样本

  • 机构审查委员会批准的单中心前瞻性随机对照试验,纳入126例接受主动脉CTA的患者,随机分配至A组(120 kVp、标准体重适配对比剂剂量)、B组(90 kVp、降低辐射剂量、标准对比剂剂量)或C组(90 kVp、标准辐射剂量、降低对比剂剂量);所有方案均在第三代双源CT扫描仪上使用自动管电压选择系统的半模式完成,A组和B组滑块水平设为11(高对比任务),C组设为3(低对比任务);以对比噪声比非劣效性检验和5点Likert量表主观图像质量评估为参考标准。

诊断标准与关键数值

  • B组和C组的对比噪声比(CNR)与A组进行非劣效性检验,A、B、C组CNR分别为30.2±7、33.4±6.7、30.5±8.9;主观图像质量采用5点Likert量表评估,三组总体主观图像质量评分分别为4.09±0.21、4.03±0.19、4.08±0.17,组间无显著差异(P=0.4)。

临床应用与阅片要点

  • 作者认为该优化可用于扩展未来的自动管电压选择算法,以将患者肾功能等临床风险因素纳入考虑。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to compare the image quality of low-kV protocols with optimized automatic tube voltage selection (ATVS) settings to reduce either radiation dose or contrast medium (CM) with that of a reference protocol for computed tomography angiography (CTA) of the thoracoabdominal aorta.

MATERIALS AND METHODS: In this institutional review board-approved, single-center, prospective randomized controlled trial, 126 patients receiving CTA of the aorta were allocated to one of three computed tomography protocols: (A) reference protocol at 120 kVp and standard weight-adapted CM dose; (B) protocol at 90 kVp, reduced radiation and standard CM dose; and (C) protocol at 90 kVp, standard radiation and reduced CM dose. All three protocols were performed on a third-generation dual-source computed tomography scanner using the semimode of the ATVS system. The image-task-dependent optimization settings of the ATVS (slider level) were adjusted to level 11 (high-contrast task) for protocols A and B and level 3 (low-contrast task) for protocol C. Radiation dose parameters were assessed. The contrast-to-noise ratios (CNRs) of protocols B and C were tested for noninferiority compared with A. Subjective image quality was assessed using a 5-point Likert scale.

RESULTS: Size-specific dose estimate was 34.3% lower for protocol B compared with A (P < 0.0001). Contrast medium was 20.2% lower for protocol C compared with A (P < 0.0001). Mean CNR in B and C was noninferior to protocol A (CNR of 30.2 ± 7, 33.4 ± 6.7, and 30.5 ± 8.9 for protocols A, B, and C, respectively). There was no significant difference in overall subjective image quality among protocols (4.09 ± 0.21, 4.03 ± 0.19, and 4.08 ± 0.17 for protocols A, B, and C, respectively; P = 0.4).

CONCLUSIONS: The slider settings of an ATVS system can be adjusted to optimize either radiation dose or CM at noninferior image quality in low-kV CTA of the aorta. This optimization could be used to extend future ATVS algorithms to take clinical risk factors like kidney function of individual patients into account.