新型C臂系统通过优化硬件与软件实现辐射剂量降低的能力
前瞻性随机双组试验纳入49例原发性或继发性肝癌行经动脉化疗栓塞的患者,比较新型C臂系统(21例)与传统血管造影系统(28例)的辐射剂量与DSA图像质量。新型系统使累积剂量面积乘积中位数降低72%(76 vs 269 Gy*cm2),透视DAP和DSA-DAP分别降低48%和77%,DSA图像质量无统计学显著差异。
要点
- 新型C臂系统使TACE累积DAP中位数降低72%(76 vs 269 Gy*cm2)。
- 透视DAP降低48%(p=0.012),DSA-DAP降低77%(p<0.01)。
- 两系统DSA图像质量无统计学显著差异。
核心结论
- 新型C臂系统显著降低TACE操作的辐射暴露,且未增加透视时间、未对DSA图像质量产生负面影响。
- 优化硬件与软件的结合带来最高辐射剂量降低,对患者和介入医师至关重要。
研究设计与样本
- 前瞻性、随机、经IRB批准的两组试验设计。
- 连续纳入49例原发性或继发性肝癌行经动脉化疗栓塞(TACE)的患者。
- 28例在传统血管造影系统B上治疗,21例在提供改进硬件和优化图像处理算法的新型系统A上治疗。
- 记录剂量面积乘积(DAP)和透视时间,DSA图像质量由两名独立盲法读者采用四级量表评估。
诊断标准与关键数值
- DSA图像质量采用四级量表由两名独立盲法读者评估。
- 两组患者特征、肿瘤负荷和透视时间无显著差异。
- 累积DAP中位数:新型系统76 Gy*cm2 vs 传统系统269 Gy*cm2,降低72%。
- 透视DAP降低48%(p=0.012),DSA-DAP降低77%(p<0.01)。
- 两成像平台间DSA图像质量无统计学显著差异。
临床应用与阅片要点
- 作者认为优化硬件与软件结合的新型C臂系统可显著降低TACE辐射暴露,且不增加透视时间或影响DSA图像质量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess the radiation dose reduction capabilities and the image quality of a new C-arm system in comparison to a standard C-arm system.
METHOD: Prospective, randomized, IRB approved two-arm trial design. 49 consecutive patients with primary or secondary liver cancer were treated with transarterial chemoembolization (TACE) on two different angiography units. 28 patients were treated on a conventional angiography unit B, 21 patients on unit A which provides improved hardware and optimized image processing algorithms. Dose area product (DAP) and fluoroscopy time were recorded. DSA image quality of all procedures was assessed on a four-rank-scale by two independent and blinded readers.
RESULTS: Both cohorts showed no significant differences with regard to patient characteristics, tumor burden and fluoroscopy time. The new system resulted in a statistically significant reduction of cumulative DAP of 72% compared to the old platform (median 76 vs. 269 Gy*cm2). Individually, Fluoro-DAP and DSA-DAP decreased by 48% and 77% (p = 0.012 and p < 0.01), respectively. No statistically significant differences in DSA image quality were found between the two imaging platforms.
CONCLUSIONS: The new C-arm system significantly reduced radiation exposure for TACE procedures without increased radiation time or negative impact on DSA image quality. The combination of optimized hardware and software yields the highest radiation dose reduction and is of utmost importance for patients and interventionalists.
引用
- Javor D, Moyses J, Loewe C, et al. Radiation dose reduction capabilities of a new C-arm system with optimized hard- and software. Eur J Radiol. 2020. PMID: 33221618. doi:10.1016/j.ejrad.2020.109367
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。