文献速递技术与安全PETMRCT

使用轻量型MRI表面线圈改善同步PET/MRI的PET图像质量

PET Image Quality Improvement for Simultaneous PET/MRI with a Lightweight MRI Surface Coil.
PubMed 文献 · Radiology · 2020-11-03

本研究通过体模和前瞻性在体研究评估轻量型前阵列表面线圈对PET/MRI图像质量的影响。体模研究显示轻量型线圈较传统线圈降低PET定量偏差(TOF -4.3% vs -9.0%,非TOF -4.0% vs -8.6%),在体研究肝脏SUV误差也降低(-2.4% vs -5.9%),结论为轻量型线圈使定量偏差降低超过50%。

要点

  • 轻量型前阵列线圈较传统线圈降低PET定量偏差超过50%。
  • 在体研究肝脏SUV误差轻量型-2.4% vs 传统-5.9%,均与无线圈参照有显著差异。
  • 使用飞行时间重建可进一步降低误差变异。

核心结论

  • 轻量型前阵列线圈较传统线圈使PET图像定量偏差降低超过50%。
  • 使用轻量型线圈和进行基于飞行时间重建均降低误差的变异。

研究设计与样本

  • 体模研究加前瞻性在体研究,使用PET/CT扫描仪在三种条件下成像:无MRI表面线圈(参照标准)、传统前阵列线圈、轻量型前阵列线圈。
  • 前阵列线圈不用于衰减校正处理以模拟临床PET/MRI。
  • 体模研究分别用和不用飞行时间重建以评估定量准确性和均匀性。
  • 在体研究纳入10名受试者(平均年龄66岁±10,6名男性),因肿瘤PET/CT检查转诊,成像时间为2019年10月至2020年2月。

诊断标准与关键数值

  • 体模研究飞行时间重建下,传统前阵列线圈平均PET定量偏差-9.0%,轻量型-4.3%。
  • 体模研究非飞行时间重建下,传统前阵列线圈平均PET定量偏差-8.6%,轻量型-4.0%。
  • 变异系数从传统线圈的4.3%(飞行时间)和6.2%(非飞行时间)降至轻量型线圈的2.1%和2.7%。
  • 在体研究肝脏标准化摄取值平均误差:传统前阵列线圈-5.9%(与无线圈比P=.002),轻量型-2.4%(与无线圈比P=.004)。

临床应用与阅片要点

  • 图像定量偏差定义为平均值的标准误,通过比较肝脏感兴趣区平均标准化摄取值与无线圈参照标准进行评估。
  • 使用Wilcoxon符号秩检验确定显著性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background During simultaneous PET/MRI, flexible MRI surface coils that lay on the patient are often omitted from PET attenuation correction processing, leading to quantification bias in PET images. Purpose To identify potential PET image quality improvement by using a recently developed lightweight MRI coil technology for the anterior array (AA) surface coil in both a phantom and in vivo study. Materials and Methods A phantom study and a prospective in vivo study were performed with a PET/CT scanner under three conditions: (a) no MRI surface coil (standard of reference), (b) traditional AA coil, and (c) lightweight AA coil. AA coils were not used in attenuation correction processing to emulate clinical PET/MRI. For the phantom study, PET images were reconstructed with and without time of flight (TOF) to assess quantification accuracy and uniformity. The in vivo study consisted of 10 participants (mean age, 66 years ± 10 [standard deviation]; six men) referred for a PET/CT oncologic examination who had undergone imaging between October 2019 and February 2020. Assessment of image quantification bias (defined as the standard error of the mean values) was conducted by comparing mean liver region of interest standardized uptake values with the no-coil standard of reference. A Wilcoxon signed-rank test was used to establish significance. Results For TOF and non-TOF, respectively, the phantom study revealed a mean PET quantification bias of -9.0% and -8.6% with the traditional AA coil and a mean PET quantification bias of -4.3% and -4.0% with the lightweight AA coil. The coefficients of variation reduced from 4.3% and 6.2% with the traditional AA coil to 2.1% and 2.7% with the lightweight AA coil, which demonstrated a homogeneity benefit from the lightweight coil that was greater with, versus without, TOF reconstruction. For the in vivo study, the mean liver standardized uptake value error was -5.9% with the traditional AA coil (P = .002 vs no coil) and -2.4% with the lightweight AA coil (P = .004 vs no coil). Conclusion The lightweight anterior array coil reduced PET image quantification bias by more than 50% compared with the traditional coil. Using the lightweight coil and performing time of flight-based reconstruction each reduced the variation of error. © RSNA, 2020 Online supplemental material is available for this article.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Deller TW, Mathew NK, Hurley SA, et al. PET Image Quality Improvement for Simultaneous PET/MRI with a Lightweight MRI Surface Coil. Radiology. 2020. PMID: 33141004. doi:10.1148/radiol.2020200967

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。