自适应统计迭代重建-V对低管电压CT冠状动脉钙化积分的影响——一项患者研究
前瞻性纳入103例冠状动脉钙化积分CT扫描患者,分别行120 kVp标准扫描及80、70 kVp低管电压扫描并应用不同水平ASIR-V重建,以120 kVp滤波反投影为参考标准。ASIR-V可降低低管电压扫描噪声,但导致钙化积分明显低估,80和70 kVp分别有29例(28.6%)和46例(44.7%)患者风险分级向更低分级偏移。
要点
- ASIR-V在低管电压下虽可降噪,但会造成冠状动脉钙化积分系统性低估。
- 80 kVp联合20% ASIR-V及70 kVp联合40% ASIR-V时噪声与参考标准相当,但风险分级改变率分别达28.6%和44.7%。
- 低管电压钙化积分重建时需注意ASIR-V水平对积分准确性的影响,风险分级可能被低估。
核心结论
- ASIR-V可降低低管电压CT扫描噪声,但与标准剂量钙化积分相比,ASIR-V引入显著不准确性并明显低估超低剂量钙化积分,即使噪声已归一化。
研究设计与样本
- 前瞻性患者研究,纳入103例行CT冠状动脉钙化积分扫描的患者,均行120 kVp标准扫描及80 kVp和70 kVp低管电压扫描,80和70 kVp扫描应用不同水平ASIR-V重建,以120 kVp滤波反投影重建为参考标准;采用新近发表的kVp适配阈值计算80和70 kVp扫描的钙化积分,并与参考标准比较;患者按钙化积分分为0、1-10、11-100、101-400、401-1000和>1000共6个风险类别。
诊断标准与关键数值
- 采用新近发表的kVp适配阈值从80和70 kVp扫描计算钙化积分;钙化积分风险分类为0、1-10、11-100、101-400、401-1000和>1000;ASIR-V水平升高导致钙化积分低估,80 kVp扫描偏倚范围为-128至-118,70 kVp扫描偏倚范围为-205至-198(与参考标准比较);80 kVp扫描20% ASIR-V和70 kVp扫描40% ASIR-V重建噪声水平与参考标准相当;从这些重建获得的钙化积分导致29例(28.6%)和46例(44.7%)患者风险分级改变,且全部向更低风险分级偏移。
临床应用与阅片要点
- 低管电压冠状动脉钙化积分扫描应用ASIR-V时,即使噪声与标准剂量相当,仍可能出现钙化积分低估和风险分级向下偏移,80 kVp联合20% ASIR-V时28.6%患者风险分级改变,70 kVp联合40% ASIR-V时44.7%患者风险分级改变。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the impact of adaptive statistical iterative reconstruction-V (ASIR-V) on the accuracy of ultra-low-dose coronary artery calcium (CAC) scoring.
MATERIALS AND METHOD: One-hundred-and-three patients who underwent computed tomography (CT) for CAC scoring were prospectively included. All underwent standard scanning with 120-kilovolt-peak (kVp) and with 80- and 70-kVp tube voltage. ASiR-V was applied to the 80- and 70-kVp scans at different levels. The 120-kVp scans reconstructed with filtered back projection served as the standard of reference. Recently published novel kVp-adapted thresholds were used for calculation of CAC scores from 80- and 70-kVp scans and the resulting CAC scores were compared against the standard of reference. Patients were stratified into six CAC score risk categories: 0, 1-10, 11-100, 101-400, 401-1000, and >1000.
RESULTS: Increasing levels of ASIR-V led to an increasing underestimation of CAC scores with bias ranging from -128 to -118 and from -205 to -198 for the 80- and 70-kVp scans, respectively, when compared with the standard of reference. Reconstruction with 20% and 40% ASIR-V for the 80- and 70-kVp scans, respectively, yielded noise levels comparable to the standard of reference. Nevertheless, a change in risk-class was observed in 29 (28.6%) and 46 (44.7%) patients, exclusively to a lower risk-class, when CAC scores were derived from these reconstructions.
CONCLUSION: ASIR-V leads to noise reduction in CT scans acquired with low tube-voltages. However, ASIR-V introduces substantial inaccuracies and marked underestimation of ultra-low-dose CAC scoring as compared with standard-dose CAC scoring despite normalization of noise.