文献速递胸部CT

纵隔窗设置下胸部CT深度学习图像重建的图像质量评估

The image quality of deep-learning image reconstruction of chest CT images on a mediastinal window setting.
PubMed 文献 · Clin Radiol · 2020-11-19

回顾性评估36例 contrast-enhanced 胸部CT薄层图像,比较FBP、ASiR-V(60%和100%)与DLIR(低、中、高)在纵隔窗(宽400 HU、窗位60 HU)的图像质量。DLIR-high客观噪声低于ASiR-V 60%且SNR、CNR更高,主观噪声、条纹伪影和整体图像质量评分最佳,但小血管评估评分差于ASiR-V 60%。

要点

  • DLIR-high在纵隔窗的噪声、条纹伪影和整体图像质量主观评分最佳。
  • DLIR-med和DLIR-high客观噪声低于ASiR-V 60%,SNR和CNR更高。
  • DLIR-med和DLIR-high小血管评估评分差于ASiR-V 60%。

关键影像表现/征象

  • 纵隔窗(宽400 HU、窗位60 HU)下评估噪声、条纹伪影、淋巴结、小血管和整体图像质量。
  • DLIR设置越高,噪声越低、SNR和CNR越高(p<0.0001)。
  • DLIR-high的图像纹理显著比ASiR-Vs更细腻(p<0.0001)。

诊断标准与关键数值

  • 与ASiR-V 60%相比,DLIR-med和DLIR-high噪声显著更低、SNR更高、CNR更高(p<0.0001)。
  • DLIR-high与ASiR-V 100%在噪声(p=0.2918)和CNR(p=0.0642)上无显著差异。
  • DLIR-high在噪声、条纹伪影和整体图像质量的主观评分最佳(p<0.0001)。
  • 与ASiR-V 60%相比,DLIR-med和DLIR-high在小血管评估上评分更差(p<0.0001)。

临床应用与阅片要点

  • 纵隔窗阅片时可关注DLIR-high在降低噪声和条纹伪影方面的优势。
  • 评估小血管时需注意DLIR-med和DLIR-high评分可能低于ASiR-V 60%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To assess the image quality of deep-learning image reconstruction (DLIR) of chest computed tomography (CT) images on a mediastinal window setting in comparison to an adaptive statistical iterative reconstruction (ASiR-V).

MATERIALS AND METHODS: Thirty-six patients were evaluated retrospectively. All patients underwent contrast-enhanced chest CT and thin-section images were reconstructed using filtered back projection (FBP); ASiR-V (60% and 100% blending setting); and DLIR (low, medium, and high settings). Image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were evaluated objectively. Two independent radiologists evaluated ASiR-V 60% and DLIR subjectively, in comparison with FBP, on a five-point scale in terms of noise, streak artefact, lymph nodes, small vessels, and overall image quality on a mediastinal window setting (width 400 HU, level 60 HU). In addition, image texture of ASiR-Vs (60% and 100%) and DLIR-high was analysed subjectively.

RESULTS: Compared with ASiR-V 60%, DLIR-med and DLIR-high showed significantly less noise, higher SNR, and higher CNR (p<0.0001). DLIR-high and ASiR-V 100% were not significantly different regarding noise (p=0.2918) and CNR (p=0.0642). At a higher DLIR setting, noise was lower and SNR and CNR were higher (p<0.0001). DLIR-high showed the best subjective scores for noise, streak artefact, and overall image quality (p<0.0001). Compared with ASiR-V 60%, DLIR-med and DLIR-high scored worse in the assessment of small vessels (p<0.0001). The image texture of DLIR-high was significantly finer than that of ASIR-Vs (p<0.0001).

CONCLUSIONS: DLIR-high improved the objective parameters and subjective image quality by reducing noise and streak artefacts and providing finer image texture.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Hata A, Yanagawa M, Yoshida Y, et al. The image quality of deep-learning image reconstruction of chest CT images on a mediastinal window setting. Clin Radiol. 2020. PMID: 33220941. doi:10.1016/j.crad.2020.10.011

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。