使用第三代双源CT扫描仪的成人囊性纤维化患者超低剂量胸部CT
本研究比较了81例成人囊性纤维化患者使用第三代双源CT与64排单源CT行低剂量胸部CT的客观与主观图像质量及辐射剂量。双源CT spiral吸气相剂量较单源CT降低25%,但总体(吸气与呼气相)剂量较单源CT增加40%;客观图像质量无显著差异,双源CT主观图像质量更优。
要点
- 双源CT螺旋吸气相剂量较单源CT降低25%,但总体(含呼气相)剂量增加40%。
- 双源CT与单源CT客观图像质量(噪声、SNR、CNR)无显著差异。
- 双源CT主观图像质量、边缘分辨率更优,条纹伪影和运动伪影更少。
关键影像表现/征象
- 双源CT图像主观整体图像质量和边缘分辨率显著优于单源CT图像。
- 双源CT图像的主观条纹伪影和运动伪影显著低于单源CT图像。
- 客观图像噪声标准差、信噪比、对比噪声比及主观图像噪声在两种扫描仪间差异无统计学意义。
诊断标准与关键数值
- 双源CT螺旋吸气相平均剂量长度乘积、容积CT剂量指数和有效辐射剂量较单源CT降低25%。
- 双源CT总体(吸气相和呼气相)平均剂量长度乘积、容积CT剂量指数和有效辐射剂量较单源CT增加40%。
临床应用与阅片要点
- 作者认为技术进步可用于降低容积扫描的辐射暴露,并可能促进螺旋呼气相采集用于小气道病变评估。
- 作者认为螺旋吸气相剂量降低可在客观图像质量相当的情况下伴随主观图像质量改善,可能允许更广泛使用该成像方式评估支气管扩张。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
INTRODUCTION: Chest computed tomography (CT) examinations are performed routinely in some cystic fibrosis (CF) centers in order to evaluate lung disease progression in CF patients. Continuous CT technological advancement in theory could allows a lower radiation exposure of CF patients during chest CT examinations without an image quality reduction, and this could become increasingly important over time in order to reduce the cumulative radiation dose effects given the continuous increase of CF patients predicted median survival.
OBJECTIVE: The aim of this study was to compare objective and subjective image quality and radiation dose between low-dose chest CT examinations performed in adult CF patients using a third-generation DSCT scanner and a 64-slices single-source CT (SSCT) scanner.
MATERIALS AND METHODS: Between January 2016 and August 2019, 81 CF patients underwent low-dose chest CT examinations using both a 64-slices SSCT scanner (2016-2017) and a third-generation DSCT scanner (2018-2019). Objective image noise standard deviation (INSD), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), overall subjective image quality (OSIQ), subjective image noise (SIN), subjective evaluation of streaking artifacts (SA), movement artifacts (MA) and edge resolution (ER), dose-length product (DLP), volume computed tomography dose index (CTDIvol) and effective radiation dose (ERD) were compared between DSCT and SSCT examinations. DSCT examinations consisted in spiral inspiratory end expiratory acquisitions. SSCT examinations consisted in spiral inspiratory acquisitions and five axial expiratory ones.
RESULTS: DSCT protocol showed statistically significant lower spiral inspiratory phase mean DLP, CTDIvol and ERD than SSCT protocol, with a 25% DLP, CTDIvol and ERD reduction. DSCT protocol showed statistically significant higher overall (inspiratory and expiratory phases) mean DLP, CTDIvol and ERD than SSCT protocol, with a 40% DLP, CTDIvol and ERD increase. Objective image quality (INSD, SNR and CNR) and SIN differences were not statistically significant, but subjective evaluation of DSCT images showed statistically significant better OSIQ and ER, as well as statistically significant lower SA and MA with respect to SSCT images.
CONCLUSIONS: To our knowledge, this is the first study evaluating chest CT image quality and radiation dose in adult CF patients using a third-generation DSCT scanner, and it showed that technological advancements could be used in order to reduce radiation exposure of volumetric examinations. The spiral inspiratory dose reduction can be obtained with concomitant improvements in subjective image quality with comparable objective quality. This will probably allow a wider use of this imaging modality in order to assess bronchiectasis and will probably foster spiral expiratory acquisition for small airways disease evaluation.