临床可用的脑自动体积测量软件:体积测量比较及方法间可靠性验证

Clinically Available Software for Automatic Brain Volumetry: Comparisons of Volume Measurements and Validation of Intermethod Reliability.
PubMed 文献 · Korean J Radiol · 2020-11-03

本研究纳入172例受试者(45例正常健康者、85例轻度认知障碍患者、42例阿尔茨海默病患者),比较NeuroQuant®与Inbrain®两种临床可用的MR体积测量软件。配对t检验显示除杏仁核外多数脑区体积差异显著,但方法间可靠性良好至优秀(0.72<r<0.96,0.83<ICC<0.98),仅苍白球可靠性差(左侧r=0.03、ICC=0.06;右侧r=-0.05、ICC=-0.09)。效应量在多数脑区较大(0.05<r<6.15),苍白球最大、小脑最小。

要点

  • 除杏仁核外,两种软件测量的多数脑区体积差异显著。
  • 方法间可靠性良好至优秀(0.72<r<0.96,0.83<ICC<0.98),仅苍白球可靠性差。
  • 苍白球效应量最大,小脑效应量最小(0.05<r<6.15)。

关键影像表现/征象

  • 配对t检验显示除杏仁核外多数脑区两种方法间体积差异显著。
  • 苍白球可靠性差:左侧r=0.03、ICC=0.06;右侧r=-0.05、ICC=-0.09。
  • 效应量在多数脑区较大(0.05<r<6.15),苍白球最大、小脑最小。

诊断标准与关键数值

  • 方法间可靠性良好至优秀:0.72<r<0.96,0.83<ICC<0.98(苍白球除外)。
  • 苍白球可靠性差:左侧r=0.03、ICC=0.06;右侧r=-0.05、ICC=-0.09。

临床应用与阅片要点

  • 临床医生使用这些商业软件时应考虑到,不同软件可能获得不同的体积测量值。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To compare two clinically available MR volumetry software, NeuroQuant® (NQ) and Inbrain® (IB), and examine the inter-method reliabilities and differences between them.

MATERIALS AND METHODS: This study included 172 subjects (age range, 55-88 years; mean age, 71.2 years), comprising 45 normal healthy subjects, 85 patients with mild cognitive impairment, and 42 patients with Alzheimer's disease. Magnetic resonance imaging scans were analyzed with IB and NQ. Mean differences were compared with the paired t test. Inter-method reliability was evaluated with Pearson's correlation coefficients and intraclass correlation coefficients (ICCs). Effect sizes were also obtained to document the standardized mean differences.

RESULTS: The paired t test showed significant volume differences in most regions except for the amygdala between the two methods. Nevertheless, inter-method measurements between IB and NQ showed good to excellent reliability (0.72 < r < 0.96, 0.83 < ICC < 0.98) except for the pallidum, which showed poor reliability (left: r = 0.03, ICC = 0.06; right: r = -0.05, ICC = -0.09). For the measurements of effect size, volume differences were large in most regions (0.05 < r < 6.15). The effect size was the largest in the pallidum and smallest in the cerebellum.

CONCLUSION: Comparisons between IB and NQ showed significantly different volume measurements with large effect sizes. However, they showed good to excellent inter-method reliability in volumetric measurements for all brain regions, with the exception of the pallidum. Clinicians using these commercial software should take into consideration that different volume measurements could be obtained depending on the software used.