自适应图像接收线圈技术在全脑成像中的应用

Application of Adaptive Image Receive Coil Technology for Whole-Brain Imaging.
PubMed 文献 · AJR Am J Roentgenol · 2020-11-25

研究用体模和15名健康成人比较16通道原型AIR头线圈与常规8通道和32通道头线圈的全脑MRI性能,由两名神经放射学专科医师按5级评分法评分。结果显示AIR线圈的信噪比、结构锐利度和整体图像质量评分优于8通道线圈但不及32通道线圈,g因子中位数总体低于8通道线圈但高于32通道线圈。

要点

  • 16通道AIR头线圈的全脑图像质量优于常规8通道头线圈,但不及常规32通道头线圈。
  • AIR线圈噪声协方差矩阵在不同受试者间表现稳定。
  • 16通道AIR线圈g因子中位数总体低于8通道线圈、高于32通道线圈。

核心结论

  • 平均而言,16通道原型AIR头线圈的性能优于常规8通道头线圈,但不及常规32通道头线圈。
  • 本研究显示新型AIR线圈技术用于脑部成像具有可行性,并为未来线圈设计改进提供依据。

研究设计与样本

  • 研究对体模和15名健康成人受试者进行成像。
  • 将16通道原型头AIR线圈与常规8通道和32通道头线圈在临床可用的MRI序列上进行比较。
  • 由两名委员会认证的神经放射学专科医师在共识阅片中对AIR线圈与8通道线圈、32通道线圈的多个类别按5级序数评分法评分。
  • 采用单侧和双侧Wilcoxon符号秩检验进行统计分析,并计算噪声协方差矩阵和几何因子(g因子)图。

关键影像表现/征象

  • 16通道原型AIR线圈的信噪比、结构锐利度和整体图像质量评分优于8通道线圈,但不及32通道线圈。
  • 噪声协方差矩阵显示AIR线圈在不同受试者间性能稳定。
  • 16通道AIR线圈的g因子中位数总体低于8通道线圈,但高于32通道线圈。

局限性

  • 本研究显示新型AIR线圈技术用于脑部成像的可行性,并为未来线圈设计改进提供依据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. The Adaptive Image Receive (AIR) radiofrequency coil is an emergent technology that is lightweight and flexible and exhibits electrical characteristics that overcome many of the limitations of traditional rigid coil designs. The purpose of this study was to apply the AIR coil for whole-brain imaging and compare the performance of a prototype AIR coil array with the performance of conventional head coils. SUBJECTS AND METHODS. A phantom and 15 healthy adult participants were imaged. A prototype 16-channel head AIR coil was compared with conventional 8-and 32-channel head coils using clinically available MRI sequences. During consensus review, two board-certified neuroradiologists graded the AIR coil compared with an 8-channel coil and a 32-channel coil on a 5-point ordinal scale in multiple categories. One- and two-sided Wilcoxon signed rank tests were performed. Noise covariance matrices and geometry factor (g-factor) maps were calculated. RESULTS. The signal-to-noise ratio, structural sharpness, and overall image quality scores of the prototype 16-channel AIR coil were better than those of the 8-channel coil but were not as good as those of the 32-channel coil. Noise covariance matrices showed stable performance of the AIR coil across participants. The median g-factors for the 16-channel AIR coil were, overall, less than those of the 8-channel coil but were greater than those of the 32-channel coil. CONCLUSION. On average, the prototype 16-channel head AIR coil outperformed a conventional 8-channel head coil but did not perform as well as a conventional 32-channel head coil. This study shows the feasibility of the novel AIR coil technology for imaging the brain and provides insight for future coil design improvements.