文献速递技术与安全神经骨肌CT

单能CT与双能CT在不同金属植入物尸体标本中金属伪影减少效果的比较

Comparison of metal artifact reduction using single-energy CT and dual-energy CT with various metallic implants in cadavers.
PubMed 文献 · Eur J Radiol · 2020-10-16

研究在6具含头颈腹盆及四肢金属植入物的尸体上,用320排CT比较标准、SEMAR和DECT三种方案的金属伪影减少效果。4名放射科医师盲法5分制评分显示,SEMAR在髋、膝、骨盆、腹部及颌面扫描中评分显著优于标准图像(3.25±0.88 vs 2.14±0.93,p<0.001),但SEMAR出现标准图像没有的新重建伪影;DECT仅在颈椎的伪影严重度和光滑结构显示上显著优于标准图像(3.50±0.50 vs 2.0±0.58,p<0.001),其余病例因噪声增加和低对比度分辨率下降,医师更偏好标准图像。所有病例中SEMAR均优于标准和DECT图像。

要点

  • SEMAR在髋、膝、骨盆、腹部及颌面扫描中金属伪影减少评分显著优于标准图像(3.25±0.88 vs 2.14±0.93,p<0.001)。
  • SEMAR可产生标准图像没有的新重建伪影,阅片时应同时审阅原始图像和SEMAR图像。
  • 解剖结构或植入物较小时,DECT可能优于SEMAR且不产生额外伪影,但需注意软组织对比度下降。

诊断标准与关键数值

  • 采用5分制评分量表评估金属伪影严重度、解剖结构显示、诊断解释和图像偏好。
  • SEMAR在髋、膝、骨盆、腹部及颌面扫描中的评分显著优于标准图像(3.25±0.88 vs 2.14±0.93,p<0.001)。
  • DECT在颈椎的金属伪影严重度和光滑结构显示评分显著优于标准图像(3.50±0.50 vs 2.0±0.58,p<0.001)。

临床应用与阅片要点

  • 放射科医师应了解SEMAR可产生新重建伪影,并同时审阅原始图像和SEMAR图像。
  • 当解剖结构或植入物相对较小时,DECT可能优于SEMAR且不产生额外伪影,但应注意软组织对比度下降。
  • 在颈椎,DECT的金属伪影严重度和光滑结构显示评分显著优于标准图像,且被1名放射科医师偏好。
  • 在其余病例中,由于DECT图像噪声增加和低对比度分辨率下降,放射科医师更偏好标准图像而非DECT图像。
  • 在所有病例中,放射科医师均偏好SEMAR图像而非标准图像和DECT图像。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The purpose of this study was to compare the effectiveness of metal artifact reduction using Single Energy Metal Artifact Reduction (SEMAR) and Dual Energy CT (DECT).

MATERIALS AND METHODS: Six cadavers containing metal implants in the head, neck, abdomen, pelvis, and extremities were scanned with Standard, SEMAR, and DECT protocols on a 320-slice CT scanner. Four specialized radiologists blinded to acquisition methods rated severity of metal artifacts, visualization of anatomic structures, diagnostic interpretation, and image preference with a 5-point grading scale.

RESULTS: Scores were significantly better for SEMAR than Standard images in the hip, knee, pelvis, abdomen, and maxillofacial scans (3.25 ± 0.88 versus 2.14 ± 0.93, p < 0.001). However, new reconstruction artifacts developed in SEMAR images that were not present in Standard images. Scores for severity of metal artifacts and visualization of smooth structures were significantly better for DECT than Standard images in the cervical spine (3.50±0.50 versus 2.0±0.58, p < 0.001) and was preferred over Standard images by one radiologist. In all other cases, radiologists preferred the Standard image over the DECT image due to increased image noise and reduced low-contrast resolution with DECT. In all cases, SEMAR was preferred over Standard and DECT images.

CONCLUSION: SEMAR was more effective at reducing metal artifacts than DECT. Radiologists should be aware of new artifacts and review both the original and SEMAR images. When the anatomy or implant is relatively small, DECT may be superior to SEMAR without additional artifacts. However, radiologist should be aware of a reduction in soft tissue contrast.

来源:Eur J Radiol,查看原文。版权归原作者所有。