文献速递骨肌CT

低剂量CT联合金属伪影减少技术在关节置换成像中的应用:尸体与临床研究

Low-dose CT with metal artifact reduction in arthroplasty imaging: a cadaveric and clinical study.
PubMed 文献 · Skeletal Radiol · 2020-10-10

本研究在2例尸体肩关节置换和25例患者肩关节置换中,比较模拟半剂量与临床剂量金属伪影减少(MAR)CT技术的可比性。结果显示模拟半剂量与全剂量在伪影评分、植入物位置测量及读者间一致性上均相近,作者认为该技术可用于降低关节置换成像的辐射剂量。

要点

  • 模拟半剂量MAR CT与标准剂量在肩关节置换评估中定量和定性均具可比性。
  • 患者中99.3%(149/150)的植入物位置测量差异在1°或1 mm以内。
  • 全剂量与模拟半剂量扫描的读者间一致率接近(骨溶解77.3% vs 76.7%)。

核心结论

  • 模拟半剂量MAR CT技术在肩关节置换评估中与标准剂量技术在定量和定性上具可比性,表明该技术可用于降低关节置换成像的剂量。

研究设计与样本

  • 研究纳入2例尸体肩关节置换和25例患者肩关节置换,临床剂量扫描参数为140 kVp、300 qrmAs,尸体另加半剂量扫描(140 kVp、150 qrmAs)。
  • 图像采用MAR CT算法在全剂量下重建,并用噪声插入算法模拟50%剂量降低;由3名盲法读者对尸体扫描的条纹伪影进行分级。
  • 患者中比较模拟半剂量与全剂量扫描的植入物位置测量差异,并比较读者对假体周围骨溶解和肌肉萎缩的分级。

关键影像表现/征象

  • 尸体中实际半剂量与模拟半剂量方法的SD平均差异为2.42 HU(95% CI [1.4, 3.4])。
  • 尸体18次比较中13次(72.2%)未见条纹伪影分级差异。
  • 患者中植入物位置测量差异在1°或1 mm以内者占149/150(99.3%)。

诊断标准与关键数值

  • 读者间一致率:全剂量骨溶解77.3%(232/300)、肌肉萎缩76.9%(173/225);模拟半剂量骨溶解76.7%(920/1200)、肌肉萎缩74.0%(666/900)。

临床应用与阅片要点

  • 作者认为模拟半剂量MAR CT技术可用于降低关节置换成像的辐射剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To determine whether a simulated low-dose metal artifact reduction (MAR) CT technique is comparable with a clinical dose MAR technique for shoulder arthroplasty evaluation.

MATERIALS AND METHODS: Two shoulder arthroplasties in cadavers and 25 shoulder arthroplasties in patients were scanned using a clinical dose (140 kVp, 300 qrmAs); cadavers were also scanned at half dose (140 kVp, 150 qrmAs). Images were reconstructed using a MAR CT algorithm at full dose and a noise-insertion algorithm simulating 50% dose reduction. For the actual and simulated half-dose cadaver scans, differences in SD for regions of interest were assessed, and streak artifact near the arthroplasty was graded by 3 blinded readers. Simulated half-dose scans were compared with full-dose scans in patients by measuring differences in implant position and by comparing readers' grades of periprosthetic osteolysis and muscle atrophy.

RESULTS: The mean difference in SD between actual and simulated half-dose methods was 2.42 HU (95% CI [1.4, 3.4]). No differences in streak artifact grades were seen in 13/18 (72.2%) comparisons in cadavers. In patients, differences in implant position measurements were within 1° or 1 mm in 149/150 (99.3%) measurements. The inter-reader agreement rates were nearly identical when readers were using full-dose (77.3% [232/300] for osteolysis and 76.9% [173/225] for muscle atrophy) and simulated half-dose (76.7% [920/1200] for osteolysis and 74.0% [666/900] for muscle atrophy) scans.

CONCLUSION: A simulated half-dose MAR CT technique is comparable both quantitatively and qualitatively with a standard-dose technique for shoulder arthroplasty evaluation, demonstrating that this technique could be used to reduce dose in arthroplasty imaging.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Subhas N, Jun BJ, Mehta PN, et al. Low-dose CT with metal artifact reduction in arthroplasty imaging: a cadaveric and clinical study. Skeletal Radiol. 2020. PMID: 33037447. doi:10.1007/s00256-020-03643-1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。