临时组织扩张器对放疗技术影响的剂量学分析
本研究利用含临时组织扩张器的三维打印乳房模体CT图像进行剂量学分析,比较三维适形放疗、野中野放疗、调强放疗和容积旋转调强放疗在是否使用填充物及是否校正伪影条件下的测量剂量与计算剂量差异。结果显示金属端口对辐射剂量的影响与治疗技术相关,调强放疗加填充物时差异最小为0.3%,作者认为剂量差异不显著,未来需评估其临床影响。
要点
- 临时组织扩张器金属端口所致剂量差异与放疗技术相关,调强放疗加填充物时最小(0.3%)。
- 三维适形放疗加填充物并校正伪影时剂量差异为6.8%(191.6 cGy vs 179.5 cGy)。
- 容积旋转调强放疗与三维适形放疗不加填充物并校正伪影时差异分别为5.8%和5.6%。
核心结论
- 临时组织扩张器内金属端口位于辐射野内时,不同治疗技术下测量与计算剂量差异均较小,作者认为差异不显著。
- 未来研究应评估该剂量差异是否影响临床结局。
研究设计与样本
- 采用含临时组织扩张器的三维打印乳房模体采集CT图像进行剂量学分析。
- 放疗剂量为180 cGy,将13张EBT3胶片贴附于临时组织扩张器上并覆盖金属端口区域。
- 治疗计划包括三维适形放疗、野中野放疗、调强放疗和容积旋转调强放疗,并考虑是否使用填充物及是否校正伪影。
- 分析各点测量平均剂量与计算剂量的差异,以确定其与放疗技术的关联。
诊断标准与关键数值
- 三维适形放疗加填充物并校正伪影时,测量与计算剂量差异为6.8%(191.6 cGy vs 179.5 cGy)。
- 容积旋转调强放疗不加填充物并校正伪影时,剂量差异为5.8%(190.2 cGy vs 179.8 cGy)。
- 三维适形放疗不加填充物并校正伪影时,剂量差异为5.6%(193.3 cGy vs 183.1 cGy)。
- 调强放疗加填充物时剂量差异最小,为0.3%(180.7 cGy vs 181.3 cGy)。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIMS: The current study aimed to evaluate the dose effect of a temporary tissue expander (TTE) according to the radiotherapy technique for breast cancer patients.
MATERIALS AND METHODS: Computed tomography images of a 3D-printed breast phantom with a TTE were acquired for dosimetric analysis. For dose measurement during 180 cGy of radiotherapy, 13 EBT3 films were attached to the TTE while including the metal port area. Treatment planning was performed for three-dimensional conformal radiotherapy (3DCRT), field-in-field radiotherapy, intensity-modulated radiotherapy (IMRT), and volumetric modulated arc therapy (VMAT) while considering whether a bolus was used and whether artifacts were corrected. The difference at each point between the measured mean and calculated doses was analyzed to determine the association with the radiotherapy techniques.
RESULTS: The effect of the metal port on the radiation dose was associated with the treatment technique. The dose difference between the measured and calculated doses was 6.8% (191.6 cGy vs. 179.5 cGy) for cases treated with 3DCRT with bolus and artifact correction. The dose difference for cases treated with VMAT and 3DCRT without bolus and with artifact correction was 5.8% (190.2 cGy vs. 179.8 cGy) and 5.6% (193.3 cGy vs. 183.1 cGy), respectively. IMRT with the bolus showed a minimum difference of 0.3% (180.7 cGy vs. 181.3 cGy).
CONCLUSION: The presence of the metal port within the TTE in radiation fields resulted to insignificant increased dose differences according to the treatment technique. Future studies should assess whether this dose difference could affect clinical outcomes.
引用
- Park SH, Kim YS, Choi J, et al. Dosimetric analysis of the effects of a temporary tissue expander on the radiotherapy technique. Radiol Med. 2020. PMID: 33025303. doi:10.1007/s11547-020-01297-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。