乳腺弹性成像技术:在同一人群中比较应变弹性成像与剪切波弹性成像。
我们的目的是结合常规超声(US)比较应变弹性成像(SE)与剪切波弹性成像(SWE)在鉴别乳腺病变中的诊断性能。共有198例患者的203个乳腺病灶使用MyLab 90和Aixplorer超声系统接受了常规超声、SE和SWE检查。…
要点
- 常规超声AUC 0.896,优于各弹性成像单项参数。
- SE参数中SEmean的AUC最高,为0.833。
- SWE参数中Emax与Esd的AUC均为0.845。
英文原文(摘要)· 点击展开
Our purpose was to compare the diagnostic performances of strain elastography (SE) and shear-wave elastography (SWE) in differentiating breast lesions by combining with conventional ultrasound (US). A total of 198 patients with 203 breast lesions underwent conventional US, SE and SWE examination using MyLab 90 and Aixplorer US systems. The SE parameters were SEscore, fat-to-lesion ratio, gland-to-lesion ratio, muscle-to-lesion ratio and SEmean, and the SWE parameters were Emax, Emean, Emin and Esd. Conventional US had the best diagnostic performance, with an area under the curve (AUC) of 0.896. Among all SE parameters, the AUCs of SEscore, fat-to-lesion ratio and SEmean were 0.802, 0.810 and 0.833. For SWE parameters, they were 0.845, 0.746 and 0.845, respectively, for Emax, Emean and Esd. When combined with US, the sensitivity and AUC of SWE seemed to be better than those of SE (96.55% vs. 93.10%, 0.958 vs. 0.947), but no statistically significant difference existed between them.
引用
- Jia W, Luo T, Dong Y, et al. Breast Elasticity Imaging Techniques: Comparison of Strain Elastography and Shear-Wave Elastography in the Same Population. Ultrasound Med Biol. 2020. PMID: 33109379. doi:10.1016/j.ultrasmedbio.2020.09.022