新型深度学习重建软件对低剂量腹部CT客观与主观图像质量的影响
该回顾性研究对27例患者的低剂量腹部CT图像分别采用迭代重建、滤波反投影和一种新型深度学习软件重建,评估客观与主观图像质量。结果显示新软件在软组织噪声较滤波反投影降低最多42%、较迭代重建降低最多27%,信噪比和对比噪声比最高,主观质量与迭代重建相当且优于滤波反投影,作者认为可能允许约20%的剂量降低。
要点
- 新型深度学习软件在低剂量腹部CT中较滤波反投影降低噪声最多42%,较迭代重建降低最多27%。
- 新软件信噪比和对比噪声比最高,主观图像质量与迭代重建相当且优于滤波反投影。
- 作者认为该软件可能允许腹部CT扫描剂量降低约20%。
核心结论
- 新型深度学习重建软件可降低图像噪声并保持图像信息,即使与迭代重建相比亦然,可能允许腹部CT成像约20%的剂量降低。
研究设计与样本
- 经机构审查委员会批准的回顾性研究,纳入27例患者(平均年龄38±12岁),腹部低剂量CT的容积CT剂量指数为2.9±1.8 mGy。
- 图像分别采用迭代重建、滤波反投影重建,并使用新型深度学习软件进行后处理。
- 在6个不同感兴趣区评估CT值、噪声、信噪比和对比噪声比,并用信噪比、峰值信噪比、均方根误差和平均绝对误差比较重建的结构差异。
关键影像表现/征象
- 三种方法在评估的感兴趣区内平均CT值差异在1 HU以内。
- 软组织图像中,新软件较滤波反投影噪声降低最多42%,较迭代重建噪声降低最多27%。
- 新软件的信噪比和对比噪声比最高。
- 迭代重建与新软件的主观图像质量相当,且均高于滤波反投影图像。
临床应用与阅片要点
- 作者认为该软件降低图像噪声并保持图像信息,可能允许腹部CT成像约20%的剂量降低。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Modern reconstruction and post-processing software aims at reducing image noise in CT images, potentially allowing for a reduction of the employed radiation exposure. This study aimed at assessing the influence of a novel deep-learning based software on the subjective and objective image quality compared to two traditional methods [filtered back-projection (FBP), iterative reconstruction (IR)].
METHODS: In this institutional review board-approved retrospective study, abdominal low-dose CT images of 27 patients (mean age 38 ± 12 years, volumetric CT dose index 2.9 ± 1.8 mGy) were reconstructed with IR, FBP and, furthermore, post-processed using a novel software. For the three reconstructions, qualitative and quantitative image quality was evaluated by means of CT numbers, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in six different ROIs. Additionally, the reconstructions were compared using SNR, peak SNR, root mean square error and mean absolute error to assess structural differences.
RESULTS: On average, CT numbers varied within 1 Hounsfield unit (HU) for the three assessed methods in the assessed ROIs. In soft tissue, image noise was up to 42% lower compared to FBP and up to 27% lower to IR when applying the novel software. Consequently, SNR and CNR were highest with the novel software. For both IR and the novel software, subjective image quality was equal but higher than the image quality of FBP-images.
CONCLUSION: The assessed software reduces image noise while maintaining image information, even in comparison to IR, allowing for a potential dose reduction of approximately 20% in abdominal CT imaging.
ADVANCES IN KNOWLEDGE: The assessed software reduces image noise by up to 27% compared to IR and 48% compared to FBP while maintaining the image information.The reduced image noise allows for a potential dose reduction of approximately 20% in abdominal imaging.