CT迭代重建与深度学习重建的噪声与锐利度比较
本研究纳入37例行下肢CT血管造影的患者,比较自适应统计迭代重建-V(ASIR-V,80%和100%混合因子)与深度学习重建TrueFidelity(低、中、高强度)在血管、肝脏和肌肉中的图像噪声与锐利度。定量分析显示TFI图像锐利度显著高于ASIR-V(p<0.001),高强度TFI的SNR和CNR与AV-100相似且显著高于AV-80,整体图像质量最佳。
要点
- 深度学习重建TrueFidelity在高强度设置下噪声与锐利度平衡最佳。
- TFI图像锐利度显著优于ASIR-V(p<0.001),模糊度量值更低。
- 迭代与深度学习重建中,混合或强度因子越高,噪声越低但锐利度越差。
关键影像表现/征象
- ASIR-V混合因子或TFI强度越高,所有结构的噪声越低,SNR和CNR值越高,模糊度量值越高。
- TFI图像的模糊度量值显著低于ASIR-V图像(p < 0.001),提示锐利度增加。
- 在所有研究的图像处理中,TF-H图像的整体定性图像质量最佳。
诊断标准与关键数值
- 定量分析指标包括CT值、图像噪声、信噪比(SNR)、对比噪声比(CNR)及模糊度量值(表示图像锐利度程度)。
- 定性分析评估量子斑驳和模糊程度。
- TF-H图像的SNR和CNR值显著高于AV-80图像,与AV-100图像相似。
临床应用与阅片要点
- 在日常下肢CTA阅片中,高强度深度学习重建(TF-H)可在噪声与锐利度间取得较好平衡。
- 选择迭代重建时需注意,提高混合因子虽降低噪声但会降低锐利度。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare image noise and sharpness of vessels, liver, and muscle in lower extremity CT angiography between "adaptive statistical iterative reconstruction-V" (ASIR-V) and deep learning reconstruction "TrueFidelity" (TFI).
METHODS: Thirty-seven patients (mean age, 65.2 years; 32 men) with lower extremity CT angiography were enrolled between November and December 2019. Images were reconstructed with two ASIR-V (blending factor of 80% and 100% (AV-100)) and three TFI (low-, medium-, and high-strength-level (TF-H) settings). Two radiologists evaluated these images for vessels (aorta, femoral artery, and popliteal artery), liver, and psoas muscle. For quantitative analyses, conventional indicators (CT number, image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR)) and blur metric values (indicating the degree of image sharpness) of selected regions of interest were determined. For qualitative analyses, the degrees of quantum mottle and blurring were assessed.
RESULTS: The higher the blending factor in ASIR-V or the strength in TFI, the lower the noise, the higher the SNR and CNR values, and the higher the blur metric values in all structures. The SNR and CNR values of TF-H images were significantly higher than those of AV-80 images and similar to those of AV-100 images. The blur metric values in TFI images were significantly lower than those in ASIR-V images (p < 0.001), indicating increased sharpness. Among all the investigated image procedures, the overall qualitative image quality was best in TF-H images.
CONCLUSION: TF-H was the most balanced image in terms of image noise and sharpness among the examined image combinations.
KEY POINTS: • Deep learning image reconstruction "TrueFidelity" is superior to iterative reconstruction "ASIR-V" regarding image noise and sharpness. • The high-strength "TrueFidelity" approach generated the best image quality among the examined image reconstruction procedures. • In iterative and deep learning CT image reconstruction, the higher the blending and strength factors, the lower the image noise and the poorer the image sharpness.
引用
- Park C, Choo KS, Jung Y, et al. CT iterative vs deep learning reconstruction: comparison of noise and sharpness. Eur Radiol. 2020. PMID: 33057781. doi:10.1007/s00330-020-07358-8
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。