文献速递心血管US

斑点追踪纵向应变:验证及扫描仪设置的影响

Speckle Tracking-Derived Longitudinal Strain: Validation and Influence of Scanner Settings.
PubMed 文献 · Ultrasound Med Biol · 2020-10-17

研究在35例日常患者中评估机器设置对整体纵向峰值收缩应变(GLPSS)的影响,发现增益和帧率不系统性改变GLPSS,而较高探头频率和较浅扇区深度与绝对GLPSS值轻微但统计学显著降低相关。GLPSS与射血分数相关性好(r=-0.778),观察者内和观察者间变异性重复性满意。

要点

  • 增益和帧率改变不系统性影响GLPSS。
  • 较高探头频率和较浅扇区深度与绝对GLPSS值轻微降低相关。
  • GLPSS与射血分数相关性好(r=-0.778)。

核心结论

  • 在应用成人超声心动图常用设置时,GLPSS分析在患者队列中可重复且稳健。

研究设计与样本

  • 研究纳入35例日常患者,每例患者四腔心切面以不同机器参数多次记录。
  • 射血分数范围10%至76%,与GLPSS相关性好(r=-0.778)。

诊断标准与关键数值

  • GLPSS未因增益和帧率修改而系统性改变。
  • 较高探头频率与绝对GLPSS值轻微但统计学显著降低相关:谐波成像频率平均效应1.102%/MHz(p < 0.001),基波成像频率平均效应0.662%/MHz(p = 0.033)。
  • 较浅扇区深度与绝对GLPSS值轻微但统计学显著降低相关:平均效应-0.156%/cm(p < 0.001)。

临床应用与阅片要点

  • 在成人超声心动图常用设置下,GLPSS分析可重复且稳健。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Speckle tracking-based strain analysis is an evolving technique for the assessment of left ventricular function. We evaluated the influence of machine settings on global longitudinal peak systolic strain (GLPSS) values in an everyday patient population (n = 35). In each patient, the four-chamber view was recorded multiple times with different machine parameters. Ejection fraction ranged between 10% and 76% and correlated well with GLPSS (r = -0.778). GLPSS was not altered systematically by modifications of gain and frame rate. Conversely, higher transducer frequencies (mean effect: 1.102%/MHz, p < 0.001, and 0.662%/MHz, p = 0.033, for harmonic and fundamental imaging frequencies, respectively) and lower sector depth (mean effect: -0.156%/cm, p < 0.001) were associated with a slight-but statistically significant-reduction in absolute GLPSS values. Intra- and inter-observer variability exhibited satisfactory repeatability. GLPSS analysis proved to be reproducible and robust in our patient cohort if common settings for adult echocardiography were applied.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Blessberger H, Bergler-Klein J, Graf S, et al. Speckle Tracking-Derived Longitudinal Strain: Validation and Influence of Scanner Settings. Ultrasound Med Biol. 2020. PMID: 33077338. doi:10.1016/j.ultrasmedbio.2020.09.005

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。