文献速递胸部CT

第三代迭代重建在双源高螺距低剂量胸部CT协议中的应用:采用锡滤过光谱整形100 kV对少量COVID-19患者的研究

Third-generation iterative reconstruction on a dual-source, high-pitch, low-dose chest CT protocol with tin filter for spectral shaping at 100 kV: a study on a small series of COVID-19 patients.
PubMed 文献 · Radiol Med · 2020-10-12

研究纳入75例确诊COVID-19住院患者,行双能量CT(DECT,90/150Sn kV)和双源高螺距锡滤过100 kVp低剂量CT(LDCT),比较滤波反投影、ADMIRE 3和ADMIRE 5重建的图像质量与肺内病变显示。LDCT中位有效剂量0.39 mSv、采集时间0.71 s,运动伪影显著少于DECT(p<0.00001);DECT3和LDCT3图像质量及肺解剖与病变显示最佳。

要点

  • LDCT中位有效剂量0.39 mSv、采集时间0.71 s,运动伪影显著少于DECT(p<0.00001)。
  • DECT3与LDCT3的图像质量、肺解剖及病变显示最佳,各系列间差异显著(p<0.00001)。
  • ADMIRE显著提高LDCT的SNR(p<0.00001),超重患者中几乎无显著差异。

核心结论

  • 采用光谱整形和ADMIRE3的LDCT在COVID-19患者评估中提供可接受的图像质量,并显著降低辐射剂量和运动伪影。

研究设计与样本

  • 研究类型为对少量COVID-19患者系列的研究,纳入75例确诊COVID-19住院患者(57男/18女,中位年龄63岁)。
  • 患者接受双能量采集(DECT,90/150Sn kV)和双源高螺距锡滤过100 kVp采集(LDCT)。
  • 在DECT3上评估肺部表现及其范围(25分评分法)。
  • 两名放射科医生以共识方式用5分量表评估LDCT0、LDCT3、LDCT5和DECT3的整体图像质量、解剖结构和基本征象。
  • 记录信噪比(SNR)、体重指数、曝光时间和辐射剂量。

关键影像表现/征象

  • 中位肺部累及范围为13/25分。
  • 影像学表现的检出比例与现有文献相当。

诊断标准与关键数值

  • 肺部累及范围采用25分评分法评估,中位得分为13/25分。
  • 图像质量、解剖结构和基本征象采用5分量表评估。
  • ADMIRE显著提高LDCT的SNR(p<0.00001),超重患者中几乎无显著差异。
  • LDCT中位有效剂量为0.39 mSv,采集时间为0.71 s,运动伪影显著少于DECT(p<0.00001)。
  • DECT3和LDCT3提供最佳图像质量及肺解剖和影像表现显示,所有系列间差异显著(p<0.00001)。

临床应用与阅片要点

  • 在COVID-19患者胸部CT评估中,采用光谱整形和ADMIRE3的LDCT可提供可接受的图像质量。
  • LDCT较DECT显著降低辐射剂量和运动伪影。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the role of third-generation iterative reconstruction (ADMIRE) in dual-source, high-pitch chest CT protocol with spectral shaping at 100 kVp in Coronavirus disease 2019 (COVID-19).

METHODS: Confirmed COVID-19 inpatients undergoing to unenhanced chest CT were scanned with a dual-energy acquisition (DECT, 90/150Sn kV) and a dual-source, high-pitch acquisition with tin-filtered 100 kVp (LDCT). On the DECT with ADMIRE 3 (DECT3) were evaluated the pulmonary findings and their extension (25-point score). Two radiologists in consensus evaluated with 5-point scales the overall image quality, the anatomical structures, and the elementary findings on LDCT reconstructed with filtered backprojection (LDCT0), with ADMIRE 3 (LDCT3) and 5 (LDCT5), and on DECT3. The signal-to-noise ratio (SNR), the body mass index, the exposure times, and the radiation doses were recorded.

RESULTS: Seventy-five patients (57 M/18F; median age: 63 y.o.) were included, with median pulmonary extension of 13/25 points. The imaging findings were detected in proportion comparable to the available literature. The ADMIRE significantly improved the SNR in LDCT (p < 0.00001) with almost no significant differences in overweight patients. The LDCT had median effective dose of 0.39 mSv and acquisition time of 0.71 s with significantly less motion artifacts than DECT (p < 0.00001). The DECT3 and LDCT3 provided the best image quality and depiction of pulmonary anatomy and imaging findings, with significant differences among all the series (p < 0.00001).

CONCLUSION: The LDCT with spectral shaping and ADMIRE3 provided acceptable image quality in the evaluation of patients with COVID-19, with significantly reduced radiation dose and motion artifacts.

来源:Radiol Med,查看原文。版权归原作者所有。

引用

  1. Agostini A, Borgheresi A, Carotti M, et al. Third-generation iterative reconstruction on a dual-source, high-pitch, low-dose chest CT protocol with tin filter for spectral shaping at 100 kV: a study on a small series of COVID-19 patients. Radiol Med. 2020. PMID: 33044732. doi:10.1007/s11547-020-01298-5