运动状态下三维打印肺肿瘤模型的虚拟平扫双能CT与真实平扫CT在放疗勾画中的比较:体模研究

Comparison of virtual non-contrast dual-energy CT and a true non-contrast CT for contouring in radiotherapy of 3D printed lung tumour models in motion: a phantom study.
PubMed 文献 · Br J Radiol · 2020-10-01

研究在呼吸运动体模中比较双源双能CT与单源序列双能CT生成的虚拟平扫图像与真实平扫CT对三维打印肺肿瘤模型大体肿瘤体积和CT值的影响,结果显示单源双能CT虚拟平扫在运动模型中无法充分去除碘,而双源双能CT虚拟平扫可能是真实平扫CT的有效替代。

要点

  • 双源双能CT虚拟平扫与真实平扫CT在运动肺肿瘤模型外轮廓GTV体积上无显著差异(p>0.05)。
  • 单源序列双能CT虚拟平扫在运动模型中无法充分从水中减去碘,内芯CT值偏差达104.4±71.2 HU。
  • 双源双能CT虚拟平扫内芯CT值偏差为25.5±58.5 HU,可能作为真实平扫CT的有效替代。

核心结论

  • 单源序列双能CT虚拟平扫图像在运动肺肿瘤模型中无法充分从水中减去碘,而双源双能CT虚拟平扫图像可能是真实平扫CT的有效替代。

研究设计与样本

  • 体模研究,使用三维打印的两个直径3.0 cm球形肺肿瘤模型,内芯分别含水或碘(浓度5 mg ml-1)。
  • 在呼吸运动体模中以15次/分、振幅1.5 cm扫描,设备包括双源双能CT、单源序列双能CT和真实平扫CT,测量CT值和体积并勾画两个GTV。

关键影像表现/征象

  • 运动模型中肺肿瘤模型外轮廓GTV体积在单源双能CT和双源双能CT虚拟平扫与真实平扫CT间偏差不显著(p>0.05)。
  • 内芯相对GTV体积和CT值偏差:单源双能CT与真实平扫CT之间为6.6±0.6%和104.4±71.2 HU,双源双能CT与真实平扫CT之间为-8.4±10.6%和25.5±58.5 HU。

诊断标准与关键数值

  • 以真实平扫CT为参考,比较虚拟平扫图像的外轮廓GTV体积偏差(p>0.05为不显著)及内芯相对GTV体积和CT值偏差。

临床应用与阅片要点

  • 双源双能CT可提供增强图像用于勾画,并提供平扫图像用于运动肺肿瘤模型的放疗治疗计划。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: This work aims to investigate whether virtual non-contrast (VNC) dual-energy CT(DECT) of contrasted lung tumours can be used as an alternative for true non-contrast (TNC) images in radiotherapy. Two DECT techniques and a TNC CT were compared and influences on gross tumour volume (GTV) volume and CT number from motion artefacts in three-dimensional printed lung tumour models (LTM) in amotion phantom were examined.

METHODS: Two spherical LTMs (diameter 3.0 cm) with different inner shapes were created in a three-dimensional printer. The inner shapes contained water or iodine (concentration 5 mg ml-1) and were scanned with a dual-source DECT (ds-DECT), single-source sequential DECT (ss-DECT) and TNC CT in a respiratory motion phantom (15 breaths/min, amplitude 1.5 cm). CT number and volume of LTMs were measured. Therefore, two GTVs were contoured.

RESULTS: Deviations in GTV volume (outer shape) of LTMs in motion for contrast-enhanced ss-DECT and ds-DECT VNC images compared to TNC images are not significant (p > 0.05). Relative GTV volume and CT number deviations (inner shapes) of LTMs in motion were 6.6 ± 0.6% and 104.4 ± 71.2 HU between ss-DECT and TNC CT and -8.4 ± 10.6% and 25.5 ± 58.5 HU between ds-DECT and TNC, respectively.

CONCLUSION: ss-DECT VNC images could not sufficiently subtract iodine from water in LTMs inmotion, whereas ds-DECT VNC images might be a valid alternative to a TNC CT.

ADVANCES IN KNOWLEDGE: ds-DECT provides a contrasted image for contouring and a non-contrasted image for radiotherapy treatment planning for LTM in motion.

来源:Br J Radiol,查看原文。版权归原作者所有。