智能气管封堵装置在膈疝胎羊模型中的安全性和有效性

Safety and efficacy of smart tracheal occlusion device in diaphragmatic hernia lamb model.
PubMed 文献 · Ultrasound Obstet Gynecol · 2021-01-01

在胎羊膈疝模型中,孕97天用Smart-TO球囊行胎儿镜下气管封堵,孕116天经3.0T磁共振非侵入性使球囊放气并经超声确认,所有球囊成功放气并自行排出;与未封堵的膈疝胎羊相比,封堵组出生时肺体重比显著升高、肺泡尺寸增大但气道复杂度降低,气管组织学改变轻微但封堵组更明显。

要点

  • 孕116天经3.0T磁共振可使Smart-TO球囊非侵入性放气,超声可即刻确认。
  • 封堵组出生时肺体重比显著高于未封堵膈疝组(1.90 vs 1.07;P=0.005)。
  • 封堵组平均线性截距高于未封堵组(47.5 vs 41.9 μm;P=0.03),终末细支气管密度低于未封堵组(1.56 vs 2.23 br/mm²;P=0.005)。

关键影像表现/征象

  • 孕116天超声确认球囊在位,随后经3.0T磁共振使球囊放气,放气后立即经超声确认。
  • 所有球囊成功放气并自行从气道排出。
  • 封堵组气管组织学改变轻微,但较未封堵组更明显(气管评分2(IQR 1-3)vs 0(IQR 0-1);P=0.03)。

诊断标准与关键数值

  • 有效性以出生时肺体重比评估:封堵组1.90(IQR 1.43-2.55)vs 未封堵组1.07(IQR 0.93-1.46),P=0.005。
  • 肺形态计量学:封堵组平均线性截距47.5(IQR 45.6-48.1)μm vs 未封堵组41.9(IQR 38.8-46.1)μm,P=0.03。
  • 封堵组平均终末细支气管密度1.56(IQR 1.0-1.81)br/mm² vs 未封堵组2.23(IQR 2.14-2.40)br/mm²,P=0.005。
  • 气管组织学评分:封堵组2(IQR 1-3)vs 未封堵组0(IQR 0-1),P=0.03。

临床应用与阅片要点

  • 在胎羊模型中,经3.0T磁共振可使Smart-TO球囊非侵入性放气。
  • 放气后超声可即刻确认球囊状态,球囊可自行从胎儿上气道排出。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To investigate the efficacy and safety of the 'smart' tracheal occlusion (Smart-TO) device in fetal lambs with diaphragmatic hernia (DH).

METHODS: DH was created in fetal lambs on gestational day 70 (term, 145 days). Fetuses were allocated to either pregnancy continuation until term (DH group) or fetoscopic endoluminal tracheal occlusion (TO), performed using the Smart-TO balloon on gestational day 97 (DH + TO group). On gestational day 116, the presence of the balloon was confirmed on ultrasound, then the ewe was walked around a 3.0-Tesla magnetic resonance scanner for balloon deflation, which was confirmed by ultrasound immediately afterwards. At term, euthanasia was performed and the fetus retrieved. Efficacy of occlusion was assessed by the lung-to-body-weight ratio (LBWR) and lung morphometry. Safety parameters included tracheal side effects assessed by morphometry and balloon location after deflation. The unoccluded DH lambs served as a comparator.

RESULTS: Six fetuses were included in the DH group and seven in the DH + TO group. All balloons deflated successfully and were expelled spontaneously from the airways. In the DH + TO group, in comparison to controls, the LBWR at birth was significantly higher (1.90 (interquartile range (IQR), 1.43-2.55) vs 1.07 (IQR, 0.93-1.46); P = 0.005), while on lung morphometry, the alveolar size was significantly increased (mean linear intercept, 47.5 (IQR, 45.6-48.1) vs 41.9 (IQR, 38.8-46.1) μm; P = 0.03); whereas airway complexity was lower (mean terminal bronchiolar density, 1.56 (IQR, 1.0-1.81) vs 2.23 (IQR, 2.14-2.40) br/mm2 ; P = 0.005). Tracheal changes on histology were minimal in both groups, but more noticeable in fetal lambs that underwent TO than in unoccluded lambs (tracheal score, 2 (IQR, 1-3) vs 0 (0-1); P = 0.03).

CONCLUSIONS: In fetal lambs with DH, TO using the Smart-TO balloon is effective and safe. Occlusion can be reversed non-invasively and the deflated intact balloon expelled spontaneously from the fetal upper airways. © 2020 International Society of Ultrasound in Obstetrics and Gynecology.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。