脑动静脉畸形彩色编码定量DSA血流特征测量的评分者间信度
本研究由4名神经放射科医生对34个脑动静脉畸形的35幅彩色编码参数DSA图像独立分析,评估MTT及供血动脉和引流静脉TTP的评分者间信度。MTT测量的组内相关系数为0.218(95%CI 0.062-0.414,P=.002),信度差;当限定评分者选择同一图像、同一主要供血动脉和同一主要引流静脉时,信度提高至中等(组内相关系数0.564,95%CI 0.367-0.717,P<.001)。
要点
- 彩色编码参数DSA测量脑动静脉畸形MTT的评分者间信度差(ICC=0.218)。
- 限定同一图像、同一供血动脉和引流静脉后信度提高至中等(ICC=0.564)。
- 信度差部分源于技术敏感性和二维成像性质,微小差异可致MTT和TTP较大变异。
诊断标准与关键数值
- MTT测量的组内相关系数为0.218(95%CI 0.062-0.414,P=.002),信度差。
- 限定评分者选择同一图像、同一主要供血动脉和同一主要引流静脉时,组内相关系数为0.564(95%CI 0.367-0.717,P<.001),信度中等。
临床应用与阅片要点
- 作者认为,通过定义标准投照、供血动脉和引流静脉进行分析,可提高彩色编码参数DSA测量MTT的信度。
局限性
- 信度差部分归因于技术的敏感性和二维性质。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Hemodynamic features of brain AVMs may portend increased hemorrhage risk. Previous studies have suggested that MTT is shorter in ruptured AVMs as assessed on quantitative color-coded parametric DSA. This study assesses the interrater reliability of MTT measurements obtained using quantitative color-coded DSA.
MATERIALS AND METHODS: Thirty-five color-coded parametric DSA images of 34 brain AVMs were analyzed by 4 neuroradiologists with experience in interventional neuroradiology. Hemodynamic features assessed included MTT of the AVM and TTP of the dominant feeding artery and draining vein. Agreement among the 4 raters was assessed using the intraclass correlation coefficient.
RESULTS: The interrater reliability among the 4 raters was poor (intraclass correlation coefficient = 0.218; 95% CI, 0.062-0.414; P value = .002) as it related to MTT assessment. When the analysis was limited to cases in which the raters selected the same image to analyze and selected the same primary feeding artery and the same primary draining vein, interrater reliability improved to fair (intraclass correlation coefficient = 0.564; 95% CI, 0.367-0.717; P < .001).
CONCLUSIONS: Interrater reliability in deriving color-coded parametric DSA measurements such as MTT is poor so minor differences among raters may result in a large variance in MTT and TTP results, partly due to the sensitivity and 2D nature of the technique. Reliability can be improved by defining a standard projection, feeding artery, and draining vein for analysis.
引用
- Narsinh KH, Mueller K, Nelson J, et al. Interrater Reliability in the Measurement of Flow Characteristics on Color-Coded Quantitative DSA of Brain AVMs. AJNR Am J Neuroradiol. 2020. PMID: 33122213. doi:10.3174/ajnr.A6846
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。