1.5T介入性MRI手术室内无限深部脑刺激电极置入的安全性:连续单机构病例系列
本文报告单机构34例在1.5T介入性MRI手术室内使用第二代定向电极行「睡眠」深部脑刺激的病例,所有患者均于头皮缝合后以完全植入电极行术后MR成像,4例另于缝合前以外置电极行MR成像;未观察到MR成像相关并发症,手术并发症率与既往系列相当,提示在遵循厂商影像指南前提下该环境下使用第二代定向电极似乎安全。
要点
- 第二代定向电极在介入性MRI环境中未见MR相关并发症。
- 手术并发症率与既往第一代全向电极系列相当。
- 遵循厂商影像指南是安全使用的条件。
核心结论
- 在介入性MR成像环境中使用第二代定向电极,在遵循厂商发布的影像指南时似乎是安全的。
- 未观察到MR成像相关并发症,手术并发症率与既往系列相当。
研究设计与样本
- 单机构连续病例系列,共34例在1.5T介入性MRI手术室内行「睡眠」深部脑刺激,使用第二代定向电极。
- 所有患者于头皮缝合后以完全植入(internalized)电极行术后MR成像,其中4例另于头皮缝合前以外置(externalized)电极行MR成像。
- 描述了手术并发症与影像数据。
关键影像表现/征象
- 原文未报告影像征象。
诊断标准与关键数值
- 原文未报告敏感度、特异度、AUC、阈值或分级标准。
临床应用与阅片要点
- 作者指出,在遵循厂商发布的影像指南时,介入性MR成像环境中使用第二代定向电极似乎是安全的。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
"Asleep" deep brain stimulation using general anesthesia and intraoperative MR imaging guidance is considered "off-label" use by current FDA guidelines but is widely used in neurosurgical practice, and excellent safety has been demonstrated using first-generation, omnidirectional electrodes. Safety data for second-generation, directional electrodes in the interventional MR imaging environment have not yet been published. Herein, we report 34 cases of asleep deep brain stimulation using second-generation, directional electrodes in an interventional MR imaging suite at a single institution. Procedural complications and imaging data are described. All patients underwent postoperative MR imaging with fully implanted ("internalized") electrodes after scalp closure; 4 patients also underwent MR imaging with "externalized" electrodes before scalp closure. No MR imaging-related complications were observed, and procedural complication rates were comparable to prior series. This suggests that the use of second-generation, directional electrodes in the interventional MR imaging environment appears to be safe when following manufacturer-published imaging guidelines.
引用
- Gravbrot N, Saranathan M, Nagae LM, et al. Safety Profile of Infinity Deep Brain Stimulation Electrode Placement in a 1.5T Interventional MRI Suite: Consecutive Single-Institution Case Series. AJNR Am J Neuroradiol. 2020. PMID: 33004341. doi:10.3174/ajnr.A6776
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。