双能CT虚拟平扫与物质特异性图像的准确性:体模研究
研究用模拟腹部的大椭圆柱状体模在IQon Spectral CT和Revolution CT两台双能CT上扫描,评估虚拟平扫CT值及碘、钙、脂肪特异性图像的定量误差。虚拟平扫对模拟软组织和对比剂较准确但骨误差大,碘定量下限随辐射剂量变化,常规剂量下0.5 mg/mL为碘检测准确性的下限阈值。
要点
- 虚拟平扫对模拟软组织和对比剂的CT值误差较小,但模拟骨CT值误差可达-90.5 HU和-98.5 HU。
- 碘特异性图像在碘浓度>0.5 mg/mL时误差较低,<0.5 mg/mL时误差明显增大。
- 碘定量误差随辐射剂量降低而升高,常规剂量下0.5 mg/mL为碘检测准确性下限。
核心结论
- 虚拟平扫对模拟软组织和对比剂而言准确性尚可,但模拟骨除外。
- 碘定量的下限取决于辐射剂量,常规剂量水平下0.5 mg/mL碘浓度为碘检测准确性的下限阈值。
研究设计与样本
- 体模研究,使用模拟腹部的大椭圆柱状体模,含已知元素组成的模拟软组织、血液和骨,以及0.2–15.0 mg/mL模拟碘浓度和0.5–5.0 mg/mL碘增强血液。
- 在两台商用双能CT扫描仪上扫描:IQon Spectral CT(Philips Healthcare)和Revolution CT with Gemstone Spectral Imaging Xtream suite(GE Healthcare)。
- 测量虚拟平扫CT值的平均绝对误差,以及碘、钙、脂肪特异性图像的平均绝对百分比误差。
关键影像表现/征象
- 虚拟平扫排除模拟骨后,CT值平均绝对误差为IQon 8.0±5.0 HU、Revolution CT 9.0±6.2 HU(p=0.61)。
- 模拟骨的CT值平均误差为IQon -90.5±111.6 HU、Revolution CT -98.5±117.8 HU(p=0.08)。
诊断标准与关键数值
- 碘特异性图像在碘浓度>0.5 mg/mL时,平均绝对百分比误差为IQon 13.7%、Revolution CT 8.3%;在<0.5 mg/mL时为IQon 150%、Revolution CT 100%。
- IQon上平均绝对百分比误差从100%辐射剂量时的16.2%升至75%剂量时的18.9%和50%剂量时的24%;Revolution CT上从100%剂量时的8.8%升至75%剂量时的11.1%和50%剂量时的17.8%。
- 常规剂量水平下,0.5 mg/mL碘浓度为碘检测准确性的下限阈值。
临床应用与阅片要点
- 解读虚拟平扫时需注意骨区域CT值误差较大,可能影响骨病变或钙化的评估。
- 使用碘特异性图像定量时,碘浓度低于0.5 mg/mL时误差明显增大,常规剂量下0.5 mg/mL为准确检测的下限阈值。
- 降低辐射剂量会增大碘定量误差,低剂量扫描时需考虑碘定量的准确性下降。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The purpose of this study was to determine the quantification accuracy of virtual unenhanced images and establish the lower limit of iodine quantification as a function of dose. MATERIALS AND METHODS. A large elliptical and cylindric phantom mimicking the patient abdomen was scanned on two commercial dual-energy CT scanners, an IQon Spectral CT (Philips Healthcare) and a Revolution CT with Gemstone Spectral Imaging Xtream suite (GE Healthcare). The phantom contained simulated soft tissue, blood, and bone with known elemental composition. It also contained simulated iodine concentrations (0.2-15.0 mg/mL) and iodine-enhanced blood (0.5-5.0 mg/mL). The mean absolute error in CT value for virtual unenhanced images and mean absolute percent error in iodine, calcium, and fat-specific images were measured. RESULTS. For virtual unenhanced images, when excluding the simulated bone, the mean absolute error in CT value was 8.0 ± 5.0 (SD) HU and 9.0 ± 6.2 HU for the IQon and the Revolution CT, respectively (p = 0.61). The mean error in CT value of the simulated bone was -90.5 ± 111.6 HU and -98.5 ± 117.8 HU on the IQon and the Revolution CT, respectively (p = 0.08). For iodine-specific images, the mean absolute percent error was 13.7% and 8.3% for the IQon and the Revolution CT, respectively, above 0.5 mg/mL iodine concentration, and 150% and 100% at less than 0.5 mg/mL iodine concentration. The mean absolute percent error increased from 16.2% at 100% radiation dose to 18.9% and 24% at 75% and 50% dose, respectively, on the IQon; and from 8.8% at 100% dose to 11.1% and 17.8% at 75% and 50%, respectively, on the Revolution CT. CONCLUSION. Virtual unenhanced images are reasonably accurate for simulated soft tissues and contrast materials, except for simulated bone. The lower limit of iodine quantification is radiation-dose dependent. For typical dose levels, 0.5 mg/mL iodine concentration is the lower threshold for iodine detection accuracy.
引用
- Li B, Pomerleau M, Gupta A, et al. Accuracy of Dual-Energy CT Virtual Unenhanced and Material-Specific Images: A Phantom Study. AJR Am J Roentgenol. 2020. PMID: 32877251. doi:10.2214/AJR.19.22372
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。