文献速递神经技术与安全CTMR

深度学习合成CT用于经颅磁共振成像引导聚焦超声干预

Transcranial MR Imaging-Guided Focused Ultrasound Interventions Using Deep Learning Synthesized CT.
PubMed 文献 · AJNR Am J Neuroradiol · 2020-09-03

本研究训练并测试U-Net神经网络,将41例受试者的颅骨超短回波时间MR图像直接转换为合成CT图像,用于经颅磁共振成像引导聚焦超声的治疗计划。深度学习合成CT与真实CT的亨斯菲尔德单位空间相关系数为0.80±0.08,平均绝对误差104.57±21.33 HU,颅骨密度比估计相关系数r=0.96,模拟的峰值靶温度和温度分布与参考CT相当。结论认为深度学习方法可简化经颅磁共振成像引导聚焦超声的工作流程。

要点

  • 深度学习可将颅骨超短TE MR图像转换为合成CT,用于颅骨密度比估计和治疗计划模拟。
  • 合成CT与真实CT的HU空间相关系数为0.80±0.08,平均绝对误差104.57±21.33 HU。
  • 合成CT估计颅骨密度比的相关系数r=0.96,模拟温度与参考CT相当。

关键影像表现/征象

  • 深度学习合成的颅骨CT图像与真实颅骨CT图像高度可比。
  • 合成CT与真实CT的HU值空间相关系数为0.80±0.08,平均绝对误差为104.57±21.33 HU,受试者水平相关系数为0.91。
  • 深度学习合成颅骨CT在颅骨密度比估计中可靠,相关系数r=0.96。
  • 模拟研究显示,基于深度学习CT的峰值靶温度和温度分布与参考CT相当。

诊断标准与关键数值

  • 深度学习合成CT与真实CT的HU值空间相关系数为0.80±0.08。
  • 深度学习合成CT与真实CT的HU值平均绝对误差为104.57±21.33 HU。
  • 深度学习合成CT与真实CT的受试者水平相关系数为0.91。
  • 深度学习合成CT估计颅骨密度比与参考CT的相关系数r=0.96。

临床应用与阅片要点

  • 深度学习合成颅骨CT可用于经颅磁共振成像引导聚焦超声的治疗计划模拟,替代额外CT扫描。
  • 合成CT的颅骨密度比估计与真实CT高度相关(r=0.96),支持其在治疗计划中的应用。
  • 基于合成CT的声学和温度模拟结果与参考CT相当,可用于预测靶区温度升高。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Transcranial MR imaging-guided focused ultrasound is a promising novel technique to treat multiple disorders and diseases. Planning for transcranial MR imaging-guided focused ultrasound requires both a CT scan for skull density estimation and treatment-planning simulation and an MR imaging for target identification. It is desirable to simplify the clinical workflow of transcranial MR imaging-guided focused ultrasound treatment planning. The purpose of this study was to examine the feasibility of deep learning techniques to convert MR imaging ultrashort TE images directly to synthetic CT of the skull images for use in transcranial MR imaging-guided focused ultrasound treatment planning.

MATERIALS AND METHODS: The U-Net neural network was trained and tested on data obtained from 41 subjects (mean age, 66.4 ± 11.0 years; 15 women). The derived neural network model was evaluated using a k-fold cross-validation method. Derived acoustic properties were verified by comparing the whole skull-density ratio from deep learning synthesized CT of the skull with the reference CT of the skull. In addition, acoustic and temperature simulations were performed using the deep learning CT to predict the target temperature rise during transcranial MR imaging-guided focused ultrasound.

RESULTS: The derived deep learning model generates synthetic CT of the skull images that are highly comparable with the true CT of the skull images. Their intensities in Hounsfield units have a spatial correlation coefficient of 0.80 ± 0.08, a mean absolute error of 104.57 ± 21.33 HU, and a subject-wise correlation coefficient of 0.91. Furthermore, deep learning CT of the skull is reliable in the skull-density ratio estimation (r = 0.96). A simulation study showed that both the peak target temperatures and temperature distribution from deep learning CT are comparable with those of the reference CT.

CONCLUSIONS: The deep learning method can be used to simplify workflow associated with transcranial MR imaging-guided focused ultrasound.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Su P, Guo S, Roys S, et al. Transcranial MR Imaging-Guided Focused Ultrasound Interventions Using Deep Learning Synthesized CT. AJNR Am J Neuroradiol. 2020. PMID: 32883668. doi:10.3174/ajnr.A6758

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。