文献速递头颈儿科影像技术与安全CT

超低管电压(70 kVp)联合Flash技术用于儿童鼻旁窦低剂量CT

Low-dose CT of paediatric paranasal sinus using an ultra-low tube voltage (70 kVp) combined with the flash technique.
PubMed 文献 · Clin Radiol · 2020-09-17

研究将80例接受鼻旁窦CT的儿童按方案分为两组:A组采用80 kVp常规螺旋模式(n=40),B组采用70 kVp联合Flash扫描模式(n=40),比较辐射剂量与图像质量。70 kVp方案的CTDIvol、DLP和有效剂量均显著低于80 kVp方案(p<0.001),图像噪声增加40.7%(p=0.113)、软组织SNR降低48.9%、骨SNR增加10.1%(p=0.176和0.227),两组整体主观图像质量分级无显著差异(p=0.15)。结论认为70 kVp联合Flash技术可在维持诊断图像质量和临床可接受噪声的同时显著降低辐射剂量。

要点

  • 儿童鼻旁窦CT采用70 kVp联合Flash技术可显著降低辐射剂量。
  • 70 kVp方案图像噪声较80 kVp增加40.7%,但差异未达统计学显著(p=0.113)。
  • 两组整体主观图像质量分级无显著差异(p=0.15)。

核心结论

  • 儿童鼻旁窦成像采用超低管电压(70 kVp)联合Flash CT技术可显著降低辐射剂量,同时维持诊断图像质量和临床可接受的图像噪声。

研究设计与样本

  • 研究纳入80例接受鼻旁窦CT的儿童患者,按不同扫描方案分为两组。
  • A组采用80 kVp方案联合常规螺旋模式(n=40),B组采用70 kVp方案联合Flash扫描模式(n=40)。
  • 评估指标包括CT剂量指数(CTDIvol)、剂量长度乘积(DLP)、有效剂量(ED),以及图像噪声、信噪比(SNR)和整体主观诊断图像质量。

关键影像表现/征象

  • 70 kVp方案图像噪声较80 kVp方案增加40.7%(p=0.113)。
  • 70 kVp方案软组织SNR较80 kVp方案降低48.9%,骨SNR较80 kVp方案增加10.1%(p=0.176和0.227)。
  • 两组整体主观图像质量分级无显著差异(p=0.15)。

诊断标准与关键数值

  • 70 kVp方案CTDIvol为0.39±0.004 mGy,80 kVp方案为1.57±0.009 mGy(p<0.001)。
  • 70 kVp方案DLP为6.31±0.52 mGy·cm,80 kVp方案为19.88±2.01 mGy·cm(p<0.001)。
  • 70 kVp方案ED为0.024±0.005 mSv,80 kVp方案为0.079±0.016 mSv(p<0.001)。

临床应用与阅片要点

  • 儿童鼻旁窦CT可考虑采用70 kVp联合Flash技术以降低辐射剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To evaluate the radiation dose and diagnostic image quality of low-dose computed tomography (CT) of the paranasal sinus in children, with acquisition at an ultra-low tube voltage (70 kVp) combined with the Flash technique.

MATERIALS AND METHODS: Eighty paediatric patients underwent CT of the paranasal sinus and were divided into two groups according to different protocols (group A: 80 kVp protocol with conventional spiral mode [n=40] and group B: 70 kVp protocol with Flash scan mode [n=40]). For each examination, the CT dose index (CTDIvol), dose-length product (DLP), and effective dose (ED) were estimated. The image noise, signal-to-noise ratio (SNR), and overall subjective diagnostic image quality were also evaluated.

RESULTS: For radiation dose, the CTDIvol (mGy), DLP (mGy·cm), and ED (mSv) values of the 70 kVp protocol were significantly lower than those of the 80 kVp protocol (CTDIvol: 1.57±0.009 versus 0.39±0.004 mGy, p<0.001; DLP: 19.88±2.01 versus 6.31±0.52 mGy·cm, p<0.001; ED: 0.079±0.016 versus 0.024±0.005 mSv, p<0.001). Compared with those of the 80-kVp protocol, the image noise increased by 40.7% (p=0.113), the SNRsoft-tissue decreased by 48.9%, and the SNRbone increased by 10.1% with the 70-kVp protocol (p=0.176 and 0.227, respectively). There was no significant difference in the overall subjective image quality grades between these two groups (p=0.15).

CONCLUSION: When imaging the paranasal sinus in children, an ultra-low tube voltage (70 kVp) combined with the Flash CT technique can reduce the radiation dose significantly while maintaining diagnostic image quality with clinically acceptable image noise.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Chi J, Ji YD, Shen L, et al. Low-dose CT of paediatric paranasal sinus using an ultra-low tube voltage (70 kVp) combined with the flash technique. Clin Radiol. 2020. PMID: 32950256. doi:10.1016/j.crad.2020.08.023

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。