剂量水平对放射科医生在数字乳腺断层合成中检测微钙化的影响:一项使用乳腺体模的观察者研究

Effect of Dose Level on Radiologists' Detection of Microcalcifications in Digital Breast Tomosynthesis: An Observer Study with Breast Phantoms.
PubMed 文献 · Acad Radiol · 2020-09-16

研究使用6个5厘米厚异质性乳腺体模,植入144个模拟微钙化簇,由6名MQSA放射科医生和1名研究员在两种临床相关剂量模式下独立阅读数字乳腺断层合成(DBT)图像。结果显示,较高剂量模式(平均腺体剂量从1.34增至2.07 mGy)提高了所有尺寸斑点微钙化的检测敏感度(两个较小尺寸显著,p<0.05),减少了13.8%的假阳性簇,提高了 conspicuity 评级和置信水平(三个较大尺寸显著,p<0.05),并将每个检出簇的平均阅读时间显著缩短13.2%(p<0.05)。

要点

  • 较高剂量DBT(平均腺体剂量2.07 mGy vs 1.34 mGy)可提高微钙化检测敏感度,尤其对小尺寸斑点。
  • 较高剂量下假阳性簇标记平均减少13.8%,放射科医生置信水平提高。
  • 较高剂量下每个检出簇的平均阅读时间显著缩短13.2%。

关键影像表现/征象

  • 较高剂量模式下,所有尺寸斑点的检测敏感度均增加,其中两个较小尺寸的敏感度增加显著(p<0.05)。
  • 较高剂量模式下,平均假阳性簇标记数减少13.8%。
  • 较高剂量模式下,所有尺寸斑点的平均可见度评级和放射科医生置信水平均更高,其中三个较大尺寸达到显著(p<0.05)。
  • 较高剂量模式下,每个检出簇的平均阅读时间显著缩短13.2%(p<0.05)。

诊断标准与关键数值

  • 较高剂量模式相比低剂量模式,5厘米体模的平均腺体剂量从1.34 mGy增至2.07 mGy。

临床应用与阅片要点

  • 在5厘米厚乳腺的DBT成像中,较高剂量模式(平均腺体剂量2.07 mGy)可能提高微钙化的检测敏感度和阅读效率,并减少假阳性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare radiologists' sensitivity, confidence level, and reading efficiency of detecting microcalcifications in digital breast tomosynthesis (DBT) at two clinically relevant dose levels.

MATERIALS AND METHODS: Six 5-cm-thick heterogeneous breast phantoms embedded with a total of 144 simulated microcalcification clusters of four speck sizes were imaged at two dose modes by a clinical DBT system. The DBT volumes at the two dose levels were read independently by six MQSA radiologists and one fellow with 1-33 years (median 12 years) of experience in a fully-crossed counter-balanced manner. The radiologist located each potential cluster and rated its conspicuity and his/her confidence that the marked location contained a cluster. The differences in the results between the two dose modes were analyzed by two-tailed paired t-test.

RESULTS: Compared to the lower-dose mode, the average glandular dose in the higher-dose mode for the 5-cm phantoms increased from 1.34 to 2.07 mGy. The detection sensitivity increased for all speck sizes and significantly for the two smaller sizes (p <0.05). An average of 13.8% fewer false positive clusters was marked. The average conspicuity rating and the radiologists' confidence level were higher for all speck sizes and reached significance (p <0.05) for the three larger sizes. The average reading time per detected cluster reduced significantly (p <0.05) by an average of 13.2%.

CONCLUSION: For a 5-cm-thick breast, an increase in average glandular dose from 1.34 to 2.07 mGy for DBT imaging increased the conspicuity of microcalcifications, improved the detection sensitivity by radiologists, increased their confidence levels, reduced false positive detections, and increased the reading efficiency.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Chan HP, Helvie MA, Klein KA, et al. Effect of Dose Level on Radiologists' Detection of Microcalcifications in Digital Breast Tomosynthesis: An Observer Study with Breast Phantoms. Acad Radiol. 2020. PMID: 32950384. doi:10.1016/j.acra.2020.07.038

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。