一项针对逾6万例患者CT扫描中重复扫描的多机构研究

A Multiinstitutional Study on Wasted CT Scans for Over 60,000 Patients.
PubMed 文献 · AJR Am J Roentgenol · 2020-09-22

该回顾性研究利用自动化重复-剔除率分析算法,基于DICOM元数据量化了2015至2018年间61,102例CT检查的重复采集率。乡村站点螺旋重叠重复率为1.4%(95%CI 1.2-1.6%),学术站点为1.1%(95%CI 1.0-1.2%);两站点在团注跟踪(11.6% vs 4.3%,p<0.001)和螺旋延伸(3.3% vs 1.8%,p<0.001)导致的重复率上存在显著差异。算法灵敏度为92%(95%CI 87-96%),阳性预测值范围为91%至99%。

要点

  • 乡村站点螺旋重叠重复率1.4%(95%CI 1.2-1.6%)高于学术站点1.1%(95%CI 1.0-1.2%)。
  • 团注跟踪导致的重复率乡村站点11.6% vs 学术站点4.3%(p<0.001)。
  • 急诊成像程序的重复率往往最高,某些方案超过9%。

核心结论

  • 多机构自动化量化CT重复率可为不必要的辐射暴露提供有用指标,并识别需要培训的技术人员。

研究设计与样本

  • 回顾性研究,纳入2015至2018年间61,102例患者检查,使用自动化重复-剔除率分析算法基于DICOM元数据识别重复采集,量化一个学术站点和一个乡村站点的重复率。

关键影像表现/征象

  • 61,102例检查中4,676次重复扫描给患者带来额外辐射剂量。
  • 螺旋重叠重复率乡村站点1.4%(95%CI 1.2-1.6%),学术站点1.1%(95%CI 1.0-1.2%)。
  • 团注跟踪导致的重复率乡村站点11.6% vs 学术站点4.3%(p<0.001)。
  • 螺旋延伸导致的重复率乡村站点3.3% vs 学术站点1.8%(p<0.001)。
  • 急诊成像程序的重复率往往最高,某些方案超过9%。

诊断标准与关键数值

  • 算法识别重复采集的灵敏度为92%(95%CI 87-96%)。
  • 阳性预测值范围为91%至99%,取决于重复成像原因和站点位置。

临床应用与阅片要点

  • 该自动化方法可确定扫描仪、技术人员、方案和适应症特异性的重复率。
  • 重复率比较可用于识别需要针对性技术人员培训的领域。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. Repeated imaging is an unnecessary source of patient radiation exposure, a detriment to patient satisfaction, and a waste of time and money. Although analysis of rates of repeated and rejected images is mandated in mammography and recommended in radiography, the available data on these rates for CT are limited. MATERIALS AND METHODS. In this retrospective study, an automated repeat-reject rate analysis algorithm was used to quantify repeat rates from 61,102 patient examinations obtained between 2015 and 2018. The algorithm used DICOM metadata to identify repeat acquisitions. We quantified rates for one academic site and one rural site. The method allows scanner-, technologist-, protocol-, and indication-specific rates to be determined. Positive predictive values and sensitivity were estimated for correctly identifying and classifying repeat acquisitions. Repeat rates were compared between sites to identify areas for targeted technologist training. RESULTS. Of 61,102 examinations, 4676 instances of repeat scanning contributed excess radiation dose to patients. Estimated helical overlap repeat rates were 1.4% (95% CI, 1.2-1.6%) for the rural site and 1.1% (95% CI, 1.0-1.2%) for the academic site. Significant differences in rates of repeat imaging required because of bolus tracking (11.6% vs 4.3%; p < 0.001) and helical extension (3.3% vs 1.8%; p < 0.001) were observed between sites. Positive predictive values ranged from 91% to 99% depending on the reason for repeat imaging and site location. Sensitivity of the algorithm was 92% (95% CI, 87-96%). Rates tended to be highest for emergent imaging procedures and exceeded 9% for certain protocols. CONCLUSION. Our multiinstitutional automated quantification of repeat rates for CT provided a useful metric for unnecessary radiation exposure and identification of technologists in need of training.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。