超声造影与磁共振成像在子宫肌瘤检出与表征中的比较

Comparison of Contrast-Enhanced Ultrasound Versus Magnetic Resonance Imaging in the Detection and Characterization of Uterine Leiomyomas.
PubMed 文献 · J Ultrasound Med · 2020-09-15

本研究为前瞻性观察性研究,纳入26例女性,比较经腹超声造影(CEUS)与磁共振(MR)在检出子宫肌瘤数目、大小、位置及强化方面的表现。共成像126个肌瘤,115个(91.3%)两种检查均检出,9个(7.1%)仅MR检出,2个(1.6%)仅CEUS检出;两种方法每例患者肌瘤数目高度相关(r=0.97;P<.001),且在肌瘤数目、大小、位置及强化方面无显著差异。作者认为经腹CEUS可能作为MR的替代方法用于子宫肌瘤治疗前评估,并可作为MR禁忌患者的首选检查。

要点

  • 经腹CEUS与MR在检出子宫肌瘤数目上高度相关(r=0.97;P<.001)。
  • 126个肌瘤中115个(91.3%)两种检查均检出,9个(7.1%)仅MR检出,2个(1.6%)仅CEUS检出。
  • 两种方法在每例患者肌瘤数目、大小、位置及强化方面无显著差异。

诊断标准与关键数值

  • 共成像126个肌瘤:115个(91.3%)CEUS与MR均检出,9个(7.1%)仅MR检出,2个(1.6%)仅CEUS检出。
  • 每例患者肌瘤数目在两种方法间高度相关(r=0.97;P<.001)。
  • 每例患者肌瘤数目、大小、位置及强化在两种方法间无显著差异。

临床应用与阅片要点

  • 经腹CEUS可检出与MR相似的子宫肌瘤数目、大小、位置及强化信息。
  • 作者认为经腹CEUS可能作为MR的替代方法用于子宫肌瘤治疗前评估。
  • 作者认为经腹CEUS可作为MR禁忌患者的首选检查。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Uterine fibroids are common findings in women with pelvic pain and abnormal uterine bleeding. The reference standard test in the pretreatment evaluation of fibroids is contrast-enhanced magnetic resonance (MR) imaging. This study compared the number, size, location, and enhancement of uterine fibroids identified by contrast-enhanced ultrasound (CEUS) and MR. The aim of this study was to demonstrate that CEUS performs similarly to MR and could be used as an alternative imaging modality.

METHODS: In this prospective observational study, 26 women underwent transabdominal CEUS and MR examinations. Blinded to the original clinical MR interpretations, 2 readers reviewed the MR and CEUS studies for each patient. The number, size, location, and enhancement of each fibroid per patient were reported by MR and CEUS. A Pearson correlation coefficient was calculated for the number of fibroids identified by each modality.

RESULTS: In total, 126 fibroids were imaged: 115 (91.3%) were observed on both examinations; 9 (7.1%) were observed by MR only; and 2 (1.6%) were observed by CEUS only. A high correlation was found between the modalities for the number of fibroids identified per patient (r = 0.97; P < .001). There was also no significant difference between the modalities for each patient in the fibroid number, size, location, or enhancement.

CONCLUSIONS: These findings suggest that transabdominal CEUS may represent an alternative to MR in pretreatment evaluation of uterine fibroids and could serve as a test of choice in patients with a contraindication to MR.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Chhabra P, Daugherty R, LeNoir AM, et al. Comparison of Contrast-Enhanced Ultrasound Versus Magnetic Resonance Imaging in the Detection and Characterization of Uterine Leiomyomas. J Ultrasound Med. 2020. PMID: 32930416. doi:10.1002/jum.15495

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。