计算机断层扫描中肾结石大小与体积:采集技术和图像重建参数的影响
研究用3D打印半仿体对47枚不同成分肾结石行64排CT扫描,比较2与10 mGy CTDI及FBP、HIR、MBIR联合不同核与降噪水平的测量差异。CT总体高估结石大小(10.0±3.1 vs 8.8±2.9 mm,p<0.05),辐射剂量与降噪水平无显著影响,重建算法与核影响相关,MBIR联合锐核与手工测量最接近。
要点
- CT测量肾结石大小较手工测量总体高估(10.0±3.1 vs 8.8±2.9 mm,p<0.05),但相关性良好(R2=0.66)。
- 辐射剂量(2与10 mGy CTDI)和降噪水平对结石测量无显著影响(p>0.05)。
- 重建算法与核影响结石大小测量,MBIR联合锐核与手工测量最接近(9.3±3.1 vs 8.8±2.9 mm,p<0.05)。
关键影像表现/征象
- CT总体结石大小较手工测量高估(10.0±3.1 vs 8.8±2.9 mm,p<0.05),相关性良好(R2=0.66)。
- 单枚结石内差异最高达40%(如Man-M 5.9 mm,CT 7.3–12.0 mm)。
- MBIR联合锐核与手工测量最接近(9.3±3.1 vs 8.8±2.9 mm,p<0.05)。
诊断标准与关键数值
- 结石大小以半自动阈值分割计算最大径和体积,参考标准为数字卡尺手工测量(Man-M)。
- 采集条件为2与10 mGy CTDI,重建采用FBP、HIR、MBIR联合不同核与降噪水平。
- 统计采用ANOVA并行多重比较校正。
临床应用与阅片要点
- 报告结石大小时应了解CT较手工测量总体高估,测量值可能偏大。
- 重建算法与核影响结石大小测量,MBIR联合锐核与参考测量最接近。
- 辐射剂量与降噪水平对结石测量无显著影响(p>0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Computed tomography (CT) is routinely used to assess suspected urolithiasis. Information obtained from CT include presence, location and size of stones, with the latter frequently determining treatment strategy. While there is consensus regarding measurements procedures of kidney stones, influence of radiation dose and reconstruction techniques on stone measurements are unknown. The purpose of this study was to systematically evaluate the influence of these technical determinants on kidney stone size measurements.
METHOD: 47 kidney stones of different composition were scanned using a 64-row-multi-detector CT in a 3D-printed, semi-anthropomorphic phantom. Reference stone sizes were measured manually with a digital caliper (Man-M). Stones were imaged with 2 and 10 mGy CTDI. Images were reconstructed using filtered-back-projection, hybrid-iterative and model-based-iterative reconstruction algorithms (FBP, HIR, MBIR) in combination with different kernels and denoising levels. All stones underwent semi-automatic, threshold-based segmentation for computation of maximum diameter and volume. Statistics were conducted using ANOVA ± correction for multiple comparisons.
RESULTS: Overall stone size as compared to manual measurements was overestimated in CT (10.0 ± 3.1 vs. 8.8 ± 2.9 mm, p < 0.05) yet showing a good correlation (R2 = 0.66). Radiation dose and denoising levels did not significantly influence measurements (p > 0.05). MBIR and sharp kernels showed closest agreement with Man-M (9.3 ± 3.1 vs. 8.8 ± 2.9 mm, p < 0.05). Differences within single stones were as high as 40 % (e.g. Man-M: 5.9 mm, CT: 7.3-12.0 mm).
CONCLUSIONS: CT-based measurements of kidney stone size appear unaffected by radiation dose and denoising technique, whereas reconstruction algorithms and kernels demonstrate a relevant impact on size measurements. Smallest differences were found using MBIR with a sharp kernel.
引用
- Reimer RP, Salem J, Merkt M, et al. Size and volume of kidney stones in computed tomography: Influence of acquisition techniques and image reconstruction parameters. Eur J Radiol. 2020. PMID: 32949914. doi:10.1016/j.ejrad.2020.109267
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。