文献速递腹部技术与安全US

二维剪切波弹性成像与瞬时弹性成像测量肝硬度的比较:一项欧洲前瞻性多中心研究结果

Comparison of liver stiffness measurements by a 2D-shear wave technique and transient elastography: results from a European prospective multi-centre study.
PubMed 文献 · Eur Radiol · 2020-09-09

一项2015至2016年在四个欧洲中心开展的前瞻性研究,纳入各种慢性肝病患者,同日行佳能二维剪切波弹性成像(2D-SWE)与瞬时弹性成像(TE)肝硬度测量。在537例两者均获可靠测量的患者中,2D-SWE与TE强相关(r=0.932),但2D-SWE倾向于低估高分期肝纤维化,AUROC为0.935至0.973,正确分类390/537(73%)患者,三次2D-SWE测量即足以获得可靠肝硬度。

要点

  • 2D-SWE与TE肝硬度测量强相关(r=0.932,ICC 0.850),但2D-SWE在高值范围成比例偏低。
  • 2D-SWE按TE阈值分层的AUROC为0.935至0.973,正确分类390/537(73%)患者。
  • 三次2D-SWE测量即足以获得可靠肝硬度测量。

关键影像表现/征象

  • 537例中TE肝硬度中位值均值10.1±9.4 kPa(范围2.4-75),2D-SWE肝硬度中位值均值9.1±6.1 kPa(范围3.6-55.7),配对t检验p<0.001。
  • 两者显著相关(Pearson r=0.932,p<0.001;ICC 0.850(0.825-0.872)),Bland-Altman偏倚0.99±4.33 kPa(95%一致性界限-9.48至+7.49),高肝硬度值处呈成比例误差。
  • 2D-SWE肝硬度值随TE分类显著升高(ANOVA p<0.001)。

诊断标准与关键数值

  • 以TE肝硬度阈值为参考,2D-SWE的AUROC为0.935±0.010(95%CI 0.910-0.954)至0.973±0.009(95%CI 0.955-0.985)。
  • 2D-SWE按TE阈值正确分类390/537(73%)患者。
  • 三次2D-SWE测量足以获得可靠肝硬度测量。

临床应用与阅片要点

  • 2D-SWE与TE肝硬度强相关,可用于基于TE阈值的患者分层,但高肝硬度值时2D-SWE成比例偏低,解读高分期纤维化需注意。
  • 作者认为2D-SWE可能用于基于TE的流程管理患者。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare liver stiffness measurement (LSM) provided by Canon 2D-shear wave elastography (2D-SWE) and transient elastography (TE), the latter being the reference method.

METHODS: Prospective study conducted in four European centres from 2015 to 2016 including patients with various chronic liver diseases who had LSMs with both 2D-SWE and TE on the same day. Median of 10 valid measurements (in kPa) was used for comparison using paired t test, Pearson correlation, intraclass correlation coefficient (ICC) and Bland-Altman plot. The ability of 2D-SWE to stratify patient according to recognised LSM-TE thresholds was assessed by ROC curve analysis.

RESULTS: Six hundred forty patients were scanned, where 593 (92.7%), 572 (89.4%) and 537 (83.9%) had reliable LSMs by TE, 2D-SWE and both combined, respectively. In the latter (n = 537, 310 [57.7%] male, mean 55.3 ± 14.8 years), median LSM-TE and LSM-2D-SWE had a mean of 10.1 ± 9.4 kPa (range 2.4-75) and 9.1 ± 6.1 kPa (range 3.6-55.7) (paired t test: p < 0.001), respectively. These were significantly correlated (Pearson r = 0.932, p < 0.001, ICC 0.850 (0.825-0.872), bias 0.99 ± 4.33 kPa [95% limits of agreement - 9.48 to + 7.49] with proportional error towards higher LSM values). LSM-2D-SWE values significantly increased with TE categories (ANOVA: p < 0.001). AUROCs ranged from 0.935 ± 0.010 (95% CI 0.910-0.954) to 0.973 ± 0.009 (95% CI 0.955-0.985), resulting in correct classification of 390/537 (73%) patients. Three 2D-SWE measurements were sufficient for reliable LSMs.

CONCLUSION: LSM using 2D-SWE correlates well with TE. It tends to underestimate higher stages of liver fibrosis but correctly classifies the majority of patients. It may be used in TE-derived algorithms to manage patients.

KEY POINTS: • Liver stiffness measurement (LSM) by 2D-shear wave elastography (2D-SWE) and transient elastography (TE) are strongly correlated. • 2D-SWE shows proportionately lower LSM values compared to TE, particularly with the higher LSM range. • Three individual measurements by 2D-SWE are sufficient to assess LSM reliably.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Ronot M, Ferraioli G, Müller HP, et al. Comparison of liver stiffness measurements by a 2D-shear wave technique and transient elastography: results from a European prospective multi-centre study. Eur Radiol. 2020. PMID: 32902745. doi:10.1007/s00330-020-07212-x

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。