文献速递心血管CT

基于管电压调整对比剂注射方案在冠状动脉CT血管成像中的评估:前瞻性VOLCANIC研究结果

Evaluation of a Tube Voltage-Tailored Contrast Medium Injection Protocol for Coronary CT Angiography: Results From the Prospective VOLCANIC Study.
PubMed 文献 · AJR Am J Roentgenol · 2020-09-22

前瞻性单中心VOLCANIC-CTA研究纳入120例转诊行冠状动脉CT血管成像的患者,采用软件支持的管电压调整对比剂注射方案。图像质量在所有患者中均可诊断,70 kV组对比剂衰减峰值最高且辐射有效剂量中位数最低(2.0 mSv),各管电压水平间信噪比、对比噪声比及主观图像质量无显著差异。

要点

  • 管电压调整对比剂注射方案在70至130 kV范围内实现个体化扫描,所有患者图像质量均可诊断。
  • 70 kV组对比剂衰减峰值最高(p < 0.001),但信噪比和对比噪声比在各管电压水平间无显著差异(p ≥ 0.085)。
  • 70 kV组辐射有效剂量中位数最低,为2.0 mSv(p < 0.001)。

核心结论

  • 所提出的管电压调整注射方案可对冠状动脉CT血管成像患者进行个体化扫描,在维持高诊断图像质量的同时显著降低对比剂用量和辐射剂量。

研究设计与样本

  • 前瞻性单中心试验(VOLCANIC-CTA),纳入120例转诊行冠状动脉CT血管成像的患者,采用软件支持的管电压调整对比剂注射方案,管电压以10 kV间隔自动选择,范围70至130 kV,碘输送率根据管电压水平调整。

诊断标准与关键数值

  • 对比剂(370 mg I/mL)用量从70 kV时的33 mL(碘输送率0.7 g I/s)至130 kV时的65 mL(碘输送率1.7 g I/s);在主动脉和冠状动脉测量衰减以计算信噪比和对比噪声比,采用5分量表评估整体图像质量;所有患者图像质量评为可诊断。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. The purpose of this study was to prospectively evaluate, using software support, the feasibility and the quantitative and qualitative image quality parameters of a tube voltage-tailored contrast medium (CM) application protocol for patient-specific injection during coronary CT angiography (CCTA). SUBJECTS AND METHODS. In the Voltage-Based Contrast Media Adaptation in Coronary Computed Tomography Angiography (VOLCANIC-CTA) study, a single-center trial, 120 patients referred for CCTA were prospectively assigned to a tube voltage-tailored CM injection protocol. Automated tube voltage levels were selected in 10-kV intervals and ranged from 70 to 130 kV, and the iodine delivery rate (IDR) was adapted to the tube voltage level using dedicated software. The administered CM volume (370 mg I/mL) ranged from 33 mL at 70 kV (IDR, 0.7 g I/s) to 65 mL at 130 kV (IDR, 1.7 g I/s). Attenuation was measured in the aorta and coronary arteries to calculate quantitative signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), and 5-point scales were used to evaluate overall image quality. Radiation metrics were also assessed and compared among the protocols. RESULTS. The mean age of the study patients was 62.5 ± 11.9 (SD) years. Image quality was rated as diagnostic in all patients. Contrast attenuation peaked at 70 kV (p < 0.001), whereas SNR and CNR parameters showed no significant differences between tube voltage levels (p ≥ 0.085). Additionally, no significant differences in subjective image quality parameters were found among the different protocols (p ≥ 0.139). The lowest radiation dose values were observed in the group assigned to the 70-kV protocol, which had a median radiation effective dose of 2.0 mSv (p < 0.001). CONCLUSION. The proposed tube voltage-tailored injection protocol allows individualized scanning of patients undergoing CCTA and significantly reduces CM and radiation dose while maintaining a high diagnostic image quality.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Martin SS, Giovagnoli DA, Abadia AF, et al. Evaluation of a Tube Voltage-Tailored Contrast Medium Injection Protocol for Coronary CT Angiography: Results From the Prospective VOLCANIC Study. AJR Am J Roentgenol. 2020. PMID: 32960669. doi:10.2214/AJR.20.22777

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。