以压力导丝测量为参考确定CTA衍生血流储备分数的理想测量点
基于ADVANCE注册研究的前瞻性评估,对253例患者365支血管比较FFRCT在狭窄远端1-2 cm、压力导丝位置及远远端测量的诊断效能,以有创FFR≤0.80为参考。狭窄远端1-2 cm处FFRCT的AUC高于远远端测量,与压力导丝位置测量相似,尤其在左前降支病变中。研究支持专家近期建议,临床决策中使用狭窄远端1-2 cm处的FFRCT而非远远端测量。
要点
- FFRCT测量点应选在狭窄远端1-2 cm处,而非远远端节段。
- 狭窄远端1-2 cm处FFRCT的AUC为0.85,高于远远端0.80,与压力导丝位置0.86相似。
- 左前降支病变中狭窄远端1-2 cm处FFRCT的AUC优于远远端测量。
核心结论
- 狭窄远端1-2 cm处测量的FFRCT诊断效能高于远远端测量,与压力导丝位置测量相似,尤其在左前降支病变中。
研究设计与样本
- 研究为前瞻性评估,数据来自ADVANCE注册研究,纳入253例患者共365支血管。
- 以有创FFR≤0.80为二分类因变量,FFRCT为解释变量,用ROC AUC评估诊断准确性。
- FFRCT分别在狭窄远端1-2 cm、压力导丝位置及远远端节段测量。
诊断标准与关键数值
- 狭窄远端1-2 cm处FFRCT的AUC为0.85(95%CI 0.80-0.88),高于远远端0.80(95%CI 0.76-0.84),p=.002。
- 压力导丝位置FFRCT的AUC为0.86(95%CI 0.81-0.89),与狭窄远端1-2 cm处相似,p=.16。
- 狭窄远端1-2 cm处 vs 远远端:敏感度87% vs 92%(p=.003),特异度73% vs 42%(p<.001),PPV 75% vs 59%(p<.001),NPV 86% vs 85%(p=.72)。
- 左前降支、左回旋支及右冠状动脉亚组中,狭窄远端1-2 cm处较压力导丝位置特异度和PPV均提高(均p<.005)。
- AUC在狭窄远端1-2 cm处高于远远端仅见于左前降支病变(p<.001),左回旋支(p=.27)和右冠状动脉(p=.91)未见差异。
临床应用与阅片要点
- 研究支持专家近期建议,临床决策中使用狭窄远端1-2 cm处测量的FFRCT,而非远远端节段测量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. For clinical decision making, it was recently recommended that values of fractional flow reserve (FFR) derived from coronary CTA (FFRCT) be measured 1-2 cm distal to the stenosis, given the potential for overestimation of ischemia when FFRCT values at far distal segments are used. Supporting data are, however, lacking. OBJECTIVE. The purpose of the present study was to evaluate the diagnostic performance of FFRCT values measured 1-2 cm distal to the stenosis and at more distal locations relative to invasive FFR values. METHODS. FFRCT and invasive FFR values for 365 vessels in 253 patients identified from the Assessing Diagnostic Value of Noninvasive FFRCT in Coronary Care (ADVANCE) registry were prospectively assessed. FFRCT values were measured 1-2 cm distal to the stenosis and at the pressure wire position and far distal segments. The diagnostic accuracy of FFRCT was assessed on the basis of the ROC AUC. The AUC of FFRCT was calculated using FFRCT as an explanatory variable and an invasive FFR of 0.80 or less as the dichotomous dependent variable. RESULTS. The AUC of FFRCT values measured 1-2 cm distal to the stenosis (0.85; 95% CI, 0.80-0.88) was higher (p = .002) than that of FFRCT values measured at far distal segments (0.80; 95% CI, 0.76-0.84) and similar (p = .16) to that of FFRCT values measured at the pressure wire position (0.86; 95% CI, 0.81-0.89). FFRCT values measured 1-2 cm distal to the stenosis and at far distal segments had sensitivity of 87% versus 92% (p = .003), specificity of 73% versus 42% (p < .001), PPV of 75% versus 59% (p < .001), and NPV of 86% versus 85% (p = .72), respectively. Subgroup analyses of lesions of the left anterior descending coronary artery, left circumflex coronary artery, and right coronary artery all showed improved specificity and PPV (all p < .005) for FFRCT values measured 1-2 cm distal to the stenosis compared with values measured at the pressure wire position. However, the AUC was higher for measurements obtained 1-2 cm distal to the stenosis versus those obtained at far distal segments, for left anterior descending coronary artery lesions (p < .001) but not for left circumflex coronary artery lesions (p = .27) or right coronary artery lesions (p = .91). CONCLUSION. The diagnostic performance of FFRCT values measured 1-2 cm distal to the stenosis was higher than that of FFRCT values measured at far distal segments and was similar to that of FFRCT values measured at the pressure wire position in evaluating ischemic status, particularly for left anterior descending coronary artery lesions. CLINICAL IMPACT. The present study supports recent recommendations from experts to use FFRCT measured 1-2 cm distal to the stenosis, rather than measurements obtained at far distal segments, in clinical decision making.
引用
- Omori H, Hara M, Sobue Y, et al. Determination of the Optimal Measurement Point for Fractional Flow Reserve Derived From CTA Using Pressure Wire Assessment as Reference. AJR Am J Roentgenol. 2020. PMID: 32876482. doi:10.2214/AJR.20.24090
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。