手持式肺超声在机械通气ICU患者中线性探头与凸阵探头的比较
该研究在19例有创通气ICU患者中进行了30次12区域肺超声检查,比较手持设备线性探头与凸阵探头评估肺通气的一致性。共获得328对图像,其中29对因线性探头图像无法评分被排除。剩余图像的整体肺超声通气评分ICC为0.74,单图像通气评分ICC为0.73,B线数量ICC为0.79。
要点
- 线性与凸阵探头对单图像肺通气评分一致性良好(ICC=0.73)。
- B线数量一致性良好(ICC=0.79),整体通气评分ICC=0.74。
- 约10%的线性探头图像无法进行通气模式评分。
核心结论
- 手持式肺超声的线性探头与凸阵探头在评估机械通气ICU患者肺通气模式时一致性良好。
- 但约10%的线性探头图像无法进行通气模式评分。
研究设计与样本
- 研究纳入预期通气超过24小时的重症监护病房患者,共19例,进行30次肺超声检查。
- 使用手持超声设备的线性探头和凸阵探头分别进行12区域肺超声检查,测定每个区域的肺超声通气评分和B线数量。
- 汇总所有区域的通气评分得到整体肺超声通气评分,采用组内相关系数(ICC)评估两种探头的一致性。
诊断标准与关键数值
- 单图像肺超声通气评分ICC=0.73(95%CI 0.67-0.78)。
- B线数量ICC=0.79(95%CI 0.72-0.83)。
- 整体肺超声通气评分ICC=0.74(95%CI 0.52-0.87)。
临床应用与阅片要点
- 国际指南未推荐肺超声评估肺通气时使用特定探头。
- 使用线性探头时需注意约10%的图像可能无法评分。
局限性
- 共328对图像中29对因线性探头图像无法评分被排除分析。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
International guidelines do not recommend a specific probe for assessment of lung aeration using lung ultrasound (LUS). The aim of this study was to assess the concordance between linear and sector array probes of a handheld ultrasound device in assessment of lung aeration in invasively ventilated intensive care unit patients. This study included intensive care unit patients who were expected to be ventilated for longer than 24 h. A 12-region LUS exam was performed with a linear and a sector array probe. In each image, the LUS aeration score and number of B-lines were determined. Adding the LUS aeration scores of all regions resulted in a global LUS aeration score. Agreement between the two probes was calculated using intra-class correlation coefficients (ICCs). A total of 30 LUS exams were performed in 19 patients, resulting in a total of 328 pairs of images. Twenty-nine pairs of images were excluded from analysis because the images from the linear probe could not be scored. ICCs calculated for the remaining images revealed good concordance the LUS aeration scores for individual images (ICC = 0.73, 95% confidence interval 0.67-0.78), number of B-lines (ICC = 0.79, 95% confidence interval 0.72-0.83) and global LUS aeration score (ICC = 0.74, 95% confidence interval 0.52-0.87). In conclusion, there is good concordance between linear and sector array probes of a handheld ultrasound device in assessment of lung aeration patterns in mechanically ventilated intensive care unit patients. However, in roughly 10% of the images acquired using the linear probe, the aeration pattern could not be scored.
引用
- Smit MR, de Vos J, Pisani L, et al. Comparison of Linear and Sector Array Probe for Handheld Lung Ultrasound in Invasively Ventilated ICU Patients. Ultrasound Med Biol. 2020. PMID: 32962892. doi:10.1016/j.ultrasmedbio.2020.08.016
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。