头颈部放疗中迭代锥形束CT重建算法减少金属伪影:体模与临床研究
本研究在体模和35例有牙科金属修复体或植入物的头颈部患者中,比较迭代锥形束CT(CBCT)重建算法与FDK-CBCT对金属伪影的减少效果。结果显示迭代CBCT在体模和临床研究中均显著降低伪影指数并改善软组织显示,主观金属伪影评分和软组织可视化评分也提高。
要点
- 迭代CBCT较FDK-CBCT显著降低金属伪影指数(临床研究:149.6 vs 70.7)。
- 迭代CBCT改善金属邻近软组织的CT值准确性(临床研究:210.3 vs 69.0 HU)。
- 迭代CBCT提高金属伪影评分(2.9 vs 1.1)和软组织可视化评分(3.6 vs 1.8)。
关键影像表现/征象
- 临床研究中,舌部近金属区的HU值FDK-CBCT为210.3,迭代CBCT为69.0。
- 临床研究中,伪影指数FDK-CBCT为149.6,迭代CBCT为70.7。
- 体模研究中,HU值FDK-CBCT为389.8,迭代CBCT为-10.3。
- 体模研究中,伪影指数FDK-CBCT为322.9,迭代CBCT为96.2。
诊断标准与关键数值
- 金属伪影评分采用5分制,临床研究中FDK-CBCT为1.1,迭代CBCT为2.9。
- 软组织可视化评分采用5分制,临床研究中FDK-CBCT为1.8,迭代CBCT为3.6。
临床应用与阅片要点
- 对于有牙科金属修复体或植入物的头颈部放疗患者,迭代CBCT重建可减少金属伪影并改善软组织显示。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To investigate whether a novel iterative cone-beam computed tomography (CBCT) reconstruction algorithm reduces metal artifacts in head and neck patient images.
METHOD: An anthropomorphic phantom and 35 patients with dental metal prostheses or implants were analyzed. All CBCT images were acquired using a TrueBeam linear accelerator and reconstructed with a Feldkamp-Davis-Kress algorithm-based CBCT (FDK-CBCT) and an iterative CBCT algorithm. The mean Hounsfield unit (HU) and standard deviation values were measured on the tongue near the metal materials and the unaffected region as reference values. The artifact index (AI) was calculated. For objective image analysis, the HU value and AI were compared between FDK-CBCT and iterative CBCT images in phantom and clinical studies. Subjective image analyses of metal artifact scores and soft tissue visualizations were conducted using a five-point scale by two reviewers in the clinical study.
RESULTS: The HU value and AI showed significant artifact reduction for the iterative CBCT than for the FDK-CBCT images (phantom study: 389.8 vs.-10.3 for HU value, 322.9 vs. 96.2 for AI, FDK-CBCT vs. iterative CBCT, respectively; clinical study: 210.3 vs. 69.0 for HU value, 149.6 vs. 70.7 for AI). The subjective scores in the clinical patient study were improved in the iterative CBCT images (metal artifact score: 1.1 vs. 2.9, FDK-CBCT vs. iterative CBCT, respectively; soft tissue visualization: 1.8 vs. 3.6).
CONCLUSIONS: The iterative CBCT reconstruction algorithm substantially reduced metal artifacts caused by dental metal prostheses and improved soft tissue visualization compared to FDK-CBCT in phantom and clinical studies.