文献速递神经MR

压缩感知敏感度编码(CS-SENSE)加速脑三维T1加权快速场回波MRI

Accelerating Brain 3D T1-Weighted Turbo Field Echo MRI Using Compressed Sensing-Sensitivity Encoding (CS-SENSE).
PubMed 文献 · Eur J Radiol · 2020-08-28

前瞻性纳入33名健康对照、10例多发性硬化(MS)和10例阿尔茨海默病(AD)患者,在3.0T MR上比较常规与CS-SENSE加速因子3、4.5、6及SENSE加速因子3、4.5的3D T1W-TFE序列。CS-SENSE因子3与常规序列视觉图像质量无显著差异,时间减少65%;CS-SENSE因子6及SENSE因子3、4.5整体图像质量显著下降。基于体素的分析显示多数脑区脑体积测量ICC高(>0.9,85.2%),提示加速技术可行。

要点

  • CS-SENSE因子3加速3D T1W-TFE序列,视觉图像质量与常规序列无显著差异。
  • CS-SENSE因子3可减少65%扫描时间。
  • CS-SENSE因子6及SENSE因子3、4.5的整体图像质量显著下降。

核心结论

  • CS-SENSE因子3是加速3D T1W-TFE的合适参数,时间减少65%且视觉图像质量保留;基于体素的分析显示多数脑区脑体积测量ICC高,提示加速技术可行。

研究设计与样本

  • 前瞻性研究,纳入33名健康对照、10例多发性硬化患者和10例阿尔茨海默病患者,在3.0T MR系统上采集常规3D T1W-TFE序列及CS-SENSE因子3、4.5、6和SENSE因子3、4.5的加速序列,由两名经验丰富的放射科医师独立进行视觉评估,并在体素水平计算定量图像质量指标和组内相关系数(ICC)。

诊断标准与关键数值

  • 常规序列与CS-SENSE因子3序列的视觉图像质量指标无显著差异。
  • CS-SENSE因子6序列及SENSE因子3、4.5序列的整体图像质量显著下降。
  • 各加速扫描与常规扫描基于体素水平的ICC值高(>0.9,85.2%)。
  • 与健康对照相比,AD和MS患者在灰质萎缩方面,不同加速序列与常规序列显示一致的结果。

临床应用与阅片要点

  • 作者认为CS-SENSE因子3是加速3D T1W-TFE的合适参数,时间减少65%且视觉图像质量保留。
  • 作者指出基于体素的分析显示多数脑区脑体积测量ICC高,提示加速技术可行。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the clinical application of the accelerated 3D T1-weighted turbo field echo (T1W-TFE) using the compressed sensing-sensitivity encoding (CS-SENSE) and identify the appropriate acceleration factor.

METHODS: 33 healthy controls (HC), 10 multiple sclerosis (MS) and 10 Alzheimer's disease (AD) patients were prospectively recruited. A conventional 3D T1W-TFE sequence and accelerated sequences with CS-SENSE factors of 3, 4.5, 6 and with SENSE factors of 3, 4.5 were acquired for all participants on a 3.0T MR system. The visual evaluation was independently assessed by two experienced radiologists. Quantitative image quality metrics and intraclass correlation coefficients (ICCs) between the conventional and the accelerated sequences were performed at the voxel level. Group comparisons were performed between HC and AD or MS patients.

RESULTS: There were no significant differences in the visual image quality metrics between conventional sequence and CS-SENSE factor of 3. The sequences with CS-SENSE factor of 6 and SENSE factors of 3, 4.5 showed significantly decreased overall image quality. The ICC values based on the voxel level of each accelerated scan and conventional scan were high (>0.9, 85.2%). For different accelerated sequences, AD and MS patients showed consistent results with the conventional sequence in gray matter atrophy when compared to HC.

CONCLUSIONS: CS-SENSE factor of 3 is the appropriate parameter to accelerate the 3D T1W-TFE (65% time reduction) with preserved visual image quality. The voxel-based analysis demonstrated high ICCs for brain volume measurements in the majority of brain regions, implying the feasibility of the accelerated technique.