剪切波弹性成像定量评估新生儿脑弹性

Quantitative Evaluation of Neonatal Brain Elasticity Using Shear Wave Elastography.
PubMed 文献 · J Ultrasound Med · 2020-09-02

前瞻性观察性单中心研究纳入55例健康足月新生儿,在出生后第2至3天用4-9 MHz多频探头行脑剪切波弹性成像,测量丘脑与放射冠弹性。丘脑平均速度1.17 m/s(95%CI 1.13-1.22),放射冠1.60 m/s(95%CI 1.57-1.64),两者差异有统计学意义(P<.001);侧别、胎龄、头围、性别、身长及分娩方式对硬度测量无显著影响。

要点

  • 健康足月新生儿丘脑剪切波速度约1.17 m/s,放射冠额叶白质约1.60 m/s,两者不同。
  • 侧别、胎龄、头围、性别、身长及分娩方式对脑硬度测量无显著影响。
  • 二维脑超声剪切波弹性成像在足月新生儿中可行,可用于建立正常参考范围。

核心结论

  • 脑超声剪切波弹性成像可行,能测量新生儿脑弹性。
  • 健康足月新生儿丘脑与额叶白质放射冠的弹性不同。
  • 掌握足月新生儿正常剪切波弹性成像范围,有助于在病理条件下开展对比研究。

研究设计与样本

  • 前瞻性观察性单中心研究,连续纳入55例健康足月新生儿,于出生后第2至3天在产科病房招募。
  • 采用4-9 MHz多频探头行脑剪切波弹性成像,取丘脑与放射冠双侧矢状面,用定量剪切波弹性成像方法测硬度。
  • 每个解剖位置获取5至15个不重叠区域的弹性图,取最居中5次测量平均值作为代表值。

关键影像表现/征象

  • 丘脑平均速度1.17 m/s(95%CI 1.13-1.22 m/s),放射冠平均速度1.60 m/s(95%CI 1.57-1.64 m/s),两者差异有统计学意义(P<.001)。

诊断标准与关键数值

  • 丘脑剪切波速度均值1.17 m/s(95%CI 1.13-1.22 m/s),放射冠1.60 m/s(95%CI 1.57-1.64 m/s),对象为健康足月新生儿。

临床应用与阅片要点

  • 脑剪切波弹性成像测量时取丘脑与放射冠双侧矢状面,每个位置取5至15个不重叠区域后取最居中5次测量平均值。
  • 判读足月新生儿脑硬度时应知晓丘脑与放射冠弹性值不同,并参照相应解剖位置的正常参考值。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To demonstrate the feasibility of 2-dimensional brain ultrasound shear wave elastography (SWE) and to define the average elasticity values of the gray and white matter in term neonates.

METHODS: This work was a prospective observational single-center study including 55 healthy term neonates consecutively recruited in the maternity ward between the second and third postnatal days. All were successfully evaluated with a cerebral SWE examination performed with a multifrequency 4-9-MHz transducer. Bilateral sagittal planes of the thalamus and corona radiata were used to measure stiffness using a quantitative SWE method. Several elastograms with 5 to 15 nonoverlapping areas were obtained from the 2 different anatomic locations. The 5 most central measurements were averaged as representative values.

RESULTS: The 55 neonates ranged from 37 to 40 weeks' gestation. The estimated mean velocity values of the thalamus (1.17 m/s; 95% confidence interval, 1.13, 1.22 m/s) and corona radiata (1.60 m/s; 95% confidence interval, 1.57, 1.64 m/s) were statistically different (P < .001). There was no significant influence of laterality, gestational age, cephalic perimeter, sex, length, or type of delivery on the stiffness measurements.

CONCLUSIONS: Brain ultrasound SWE is feasible and allows measurements of neonatal brain elasticity. The elasticity of the thalamus and corona radiata at the frontal white matter in healthy term neonates is different. The knowledge of normal SWE ranges in term neonates allows comparative studies under pathologic conditions.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Garcés Iñigo E, Llorens Salvador R, Escrig R, et al. Quantitative Evaluation of Neonatal Brain Elasticity Using Shear Wave Elastography. J Ultrasound Med. 2020. PMID: 32876366. doi:10.1002/jum.15464

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。