文献速递神经头颈MR

针对高频域去噪的深度学习重建初步研究:在高分辨率三维磁共振脑池造影评估桥小脑角中的效用

A preliminary study of deep learning-based reconstruction specialized for denoising in high-frequency domain: usefulness in high-resolution three-dimensional magnetic resonance cisternography of the cerebellopontine angle.
PubMed 文献 · Neuroradiology · 2020-08-13

回顾性纳入35例接受高分辨率磁共振脑池造影的患者,图像分别用或不用深度学习重建(DLR)重建。DLR组脑脊液和脑桥的信噪比显著更高,噪声质量和整体图像质量评分显著更优,而对比度、正常侧三叉神经锐利度和伪影无显著差异。DLR通过降低图像噪声改善高分辨率磁共振脑池造影图像质量,且不牺牲对比度或锐利度。

要点

  • DLR可显著提高高分辨率磁共振脑池造影中脑脊液和脑桥的信噪比。
  • DLR不降低脑脊液与脑桥的对比度及正常侧三叉神经锐利度(FWHM)。
  • DLR显著改善噪声质量和整体图像质量,锐利度和伪影无显著差异。

核心结论

  • DLR可通过降低图像噪声改善高分辨率磁共振脑池造影的图像质量,且不牺牲对比度或锐利度。

研究设计与样本

  • 回顾性研究,纳入35例接受高分辨率磁共振脑池造影的患者,图像分别用或不用DLR重建,比较信噪比、对比度、锐利度(FWHM)及主观质量评分。

关键影像表现/征象

  • DLR组脑脊液信噪比21.81 ± 7.60 vs 无DLR组15.33 ± 4.03(p < 0.001),DLR组更高。
  • DLR组脑桥信噪比5.96 ± 1.38 vs 无DLR组3.99 ± 0.48(p < 0.001),DLR组更高。
  • 脑脊液与脑桥对比度两组无显著差异(p = 0.225)。
  • 正常侧三叉神经锐利度(FWHM)两组无显著差异(p = 0.185)。
  • 噪声质量评分DLR组3.95 ± 0.19 vs 无DLR组2.53 ± 0.44(p < 0.001),整体图像质量评分DLR组3.97 ± 0.17 vs 无DLR组2.97 ± 0.12(p < 0.001),DLR组均更高;锐利度(p = 0.371)和伪影(p = 1)两组无显著差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Deep learning-based reconstruction (DLR) has been developed to reduce image noise and increase the signal-to-noise ratio (SNR). We aimed to evaluate the efficacy of DLR for high spatial resolution (HR)-MR cisternography.

METHODS: This retrospective study included 35 patients who underwent HR-MR cisternography. The images were reconstructed with or without DLR. The SNRs of the CSF and pons, contrast of the CSF and pons, and sharpness of the normal-side trigeminal nerve using full width at half maximum (FWHM) were compared between the two image types. Noise quality, sharpness, artifacts, and overall image quality of these two types of images were qualitatively scored.

RESULTS: The SNRs of the CSF and pons were significantly higher with DLR than without DLR (CSF 21.81 ± 7.60 vs. 15.33 ± 4.03, p < 0.001; pons 5.96 ± 1.38 vs. 3.99 ± 0.48, p < 0.001). There were no significant differences in the contrast of the CSF and pons (p = 0.225) and sharpness of the normal-side trigeminal nerve using FWHM (p = 0.185) without and with DLR, respectively. Noise quality and the overall image quality were significantly higher with DLR than without DLR (noise quality 3.95 ± 0.19 vs. 2.53 ± 0.44, p < 0.001; overall image quality 3.97 ± 0.17 vs. 2.97 ± 0.12, p < 0.001). There were no significant differences in sharpness (p = 0.371) and artifacts (p = 1) without and with DLR.

CONCLUSION: DLR can improve the image quality of HR-MR cisternography by reducing image noise without sacrificing contrast or sharpness.

来源:Neuroradiology,查看原文。版权归原作者所有。

引用

  1. Uetani H, Nakaura T, Kitajima M, et al. A preliminary study of deep learning-based reconstruction specialized for denoising in high-frequency domain: usefulness in high-resolution three-dimensional magnetic resonance cisternography of the cerebellopontine angle. Neuroradiology. 2020. PMID: 32794075. doi:10.1007/s00234-020-02513-w

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。