牙科CT剂量降低:以锡滤过技术为重点的体模研究

Dose Reduction in Dental CT: A Phantom Study With Special Focus on Tin Filter Technique.
PubMed 文献 · AJR Am J Roentgenol · 2020-08-12

该体模研究用热释光探测器在20个测量点比较8种有/无锡滤过的CT协议与标准CBCT的剂量,有效全身剂量20.7–505.9 μSv,CBCT为229.3 μSv;锡滤过协议在剂量、客观图像质量和临床评价上表现最佳,高剂量协议仅在特定创伤和肿瘤问题中必要。

要点

  • 锡滤过协议在剂量、客观图像质量和临床评价上综合表现最佳。
  • 高剂量协议仅在需详细骨结构显示的创伤和肿瘤问题中必要。
  • CBCT有效剂量229.3 μSv,处于所比较方法的中上水平。

关键影像表现/征象

  • 9种检查的剂量长度乘积为5.0–111.9 mGy·cm,计算的有效全身剂量为20.7–505.9 μSv。
  • CBCT的有效剂量为229.3 μSv,处于所比较方法的中上水平。

诊断标准与关键数值

  • 基于剂量、客观图像质量和临床评价结果,锡滤过协议表现最佳。
  • 较高剂量协议在品质因数比较中显著较差,但因能详细显示骨结构,对回答某些创伤和肿瘤问题特别必要。

临床应用与阅片要点

  • 牙科CT检查可考虑使用锡滤过以降低剂量并保持良好图像质量。
  • 高剂量协议仅在特定创伤和肿瘤问题中必要。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. The purpose of this study was to determine in a phantom the dose exposure of different dental 3D sectional imaging methods (CT and cone-beam CT [CBCT]) and different CT protocols. The aim was to establish optimal protocols with the lowest possible dose and diagnostically high image quality with special consideration given to tin prefiltration. MATERIALS AND METHODS. Dose was determined with thermoluminescence detectors at 20 different measuring points on an anthropomorphic phantom. Eight different CT protocols with and without tin filtering were compared with iterative reconstruction methods and a standard CBCT protocol. Objective and subjective image evaluations and a figure-of-merit analysis of the image data were performed by radiologists and maxillofacial surgeons. RESULTS. The determined dose-length products of the nine examinations were 5.0-111.9 mGy · cm with a calculated effective whole body dose of 20.7-505.9 μSv. Cone-beam CT was in the upper midfield with an effective dose of 229.3 μSv. On the basis of dose, objective image quality, and clinical evaluation results, tin filter protocols performed best. Protocols with higher doses were significantly less useful in the figure of merit comparison but because of their detailed bony representation are particularly necessary to answer certain questions about trauma and tumors. CONCLUSION. The use of tin filtering can reduce dose in dental CT examinations, compared with standard low-dose examinations, while maintaining good image quality. The dose performance is significantly inferior even to that of a cone-beam CT examination. High-dose protocols are necessary only for certain questions.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Hackenbroch C, Schnaidt S, Halt D, et al. Dose Reduction in Dental CT: A Phantom Study With Special Focus on Tin Filter Technique. AJR Am J Roentgenol. 2020. PMID: 32783561. doi:10.2214/AJR.19.22461

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。