婴儿超高场MR成像的引入:准备与可行性
该前瞻性先导研究在12例临床稳定婴儿中于标准3T MR后立即行7T MR,模拟显示婴儿模型比吸收率未超过成人模型,新型隔音罩保证听力保护,未发生MR相关不良事件;7T下MRA、MRV、SWI、单次激发T2WI和MR波谱可获得良好质量图像,T1WI质量较低,结论为7T MR在婴儿中可行。
要点
- 婴儿7T MR扫描未发生MR相关不良事件,听力保护由新型隔音罩保证。
- 7T下MRA、MRV、SWI、单次激发T2WI和MR波谱可获得良好质量图像。
- 7T下T1WI图像质量较低。
核心结论
- 7T MR成像在婴儿中是可行的,可获得良好质量的扫描图像。
研究设计与样本
- 前瞻性先导研究,纳入临床稳定、处于足月等效年龄至矫正年龄3个月的婴儿,在标准3T MR后立即行7T MR。
- 共扫描12例婴儿,扫描过程中开发和优化7T扫描方案。
- 使用Sim4life在婴儿和成人模型中模拟7T下的比吸收率水平。
- 使用新开发的隔音罩保证听力保护,测量其噪声衰减并与3T Nordell隔音罩及无隔音罩比较。
诊断标准与关键数值
- 婴儿模型在居中位置和50 mm足方向上的整体与峰值比吸收率水平未超过成人模型。
- 新型隔音罩可保证听力保护。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Cerebral MR imaging in infants is usually performed with a field strength of up to 3T. In adults, a growing number of studies have shown added diagnostic value of 7T MR imaging. 7T MR imaging might be of additional value in infants with unexplained seizures, for example. The aim of this study was to investigate the feasibility of 7T MR imaging in infants. We provide information about the safety preparations and show the first MR images of infants at 7T.
MATERIALS AND METHODS: Specific absorption rate levels during 7T were simulated in Sim4life using infant and adult models. A newly developed acoustic hood was used to guarantee hearing protection. Acoustic noise damping of this hood was measured and compared with the 3T Nordell hood and no hood. In this prospective pilot study, clinically stable infants, between term-equivalent age and the corrected age of 3 months, underwent 7T MR imaging immediately after their standard 3T MR imaging. The 7T scan protocols were developed and optimized while scanning this cohort.
RESULTS: Global and peak specific absorption rate levels in the infant model in the centered position and 50-mm feet direction did not exceed the levels in the adult model. Hearing protection was guaranteed with the new hood. Twelve infants were scanned. No MR imaging-related adverse events occurred. It was feasible to obtain good-quality imaging at 7T for MRA, MRV, SWI, single-shot T2WI, and MR spectroscopy. T1WI had lower quality at 7T.
CONCLUSIONS: 7T MR imaging is feasible in infants, and good-quality scans could be obtained.