使用计算机模拟优化数字X线摄影系统对平均体型患者腹部、骨盆和脊柱成像的管电压范围(kVp)
本研究通过计算机模拟,让5名评估者对20例患者的腹部、骨盆和腰椎DR图像在60–120 kVp(每10 kVp递增)下按临床图像标准盲法评分,并以特定参考kVp重建的同一患者图像为对照。结果显示图像评分取决于管电压和解剖部位;腹部和骨盆的最佳管电压为70 kVp(最佳范围70–120 kVp),腰椎为100 kVp(最佳范围80–120 kVp)。
要点
- 腹部和骨盆DR成像的最佳管电压为70 kVp,最佳范围70–120 kVp。
- 腰椎DR成像的最佳管电压为100 kVp,最佳范围80–120 kVp。
- 60 kVp时所有解剖部位的所有标准评分均很差。
关键影像表现/征象
- 图像评分取决于管电压和解剖部位,两者均具有统计学显著性。
- 60 kVp时所有解剖部位的所有标准评分均很差。
- 腹部、骨盆和脊柱成像的评分分别在70、70和100 kVp时达到峰值,但其他kVp值的评分未显著更差。
临床应用与阅片要点
- 对于腹部和骨盆DR成像,可考虑使用70 kVp(最佳范围70–120 kVp)。
- 对于腰椎DR成像,可考虑使用100 kVp(最佳范围80–120 kVp)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate via computer simulation, an optimised tube voltage (kVp) range for caesium iodide (CsI)-based digital radiography (DR) of the abdomen, pelvis and lumbar spine.
METHODS: Software capable of simulating abdomen, pelvis and spine radiographs was used. Five evaluators graded clinical image criteria in images of 20 patients at tube voltages ranging from 60 to 120 kVp in 10 kVp increments. These criteria were scored blindly against the same patient reconstructed at a specific reference kVp. Linear mixed effects analysis was used to evaluate image scores for each criterion and test for statistical significance.
RESULTS: Score was dependent on tube voltage and image criteria; both were statistically significant. All criteria for all anatomies scored very poorly at 60 kVp. Scores for abdomen, pelvis and spine imaging peaked at 70, 70 and 100 kVp, respectively, but other kVp values were not significantly poorer.
CONCLUSIONS: Results indicate optimum tube voltages of 70 kVp for abdomen and pelvis (with an optimum range 70-120 kVp), and 100 kVp (optimum range 80-120 kVp) for lumbar spine.
ADVANCES IN KNOWLEDGE: There are no recommendations for optimised tube voltage parameters for DR abdomen, pelvis or lumbar spine imaging. This study has investigated and recommended an optimal tube voltage range.