文献速递腹部技术与安全CT

基于活体猪模型与拟人模体的亚毫希沃特腹部CT协议建立

Establishment of Submillisievert Abdominal CT Protocols With an In Vivo Swine Model and an Anthropomorphic Phantom.
PubMed 文献 · AJR Am J Roentgenol · 2020-07-22

研究在第三代双源CT上利用活体猪模型和拟人模体建立四种腹部低剂量CT协议,并由两名有经验放射科医师判定为可诊断。四种协议图像质量指标无显著差异,模体测量的有效剂量均不超过0.33 mSv,其中双源协议辐射剂量最低。

要点

  • 四种腹部低剂量CT协议在活体猪模型上均被两名有经验放射科医师判定为可诊断。
  • 四种低剂量协议的信噪比和对比噪声比无统计学显著差异。
  • 模体测量显示所有协议有效剂量均不超过0.33 mSv,双源协议辐射剂量最低。

关键影像表现/征象

  • 四种协议的可诊断性、空间分辨率和噪声均被评为最佳,因此均被视为可诊断。
  • 四种低剂量协议在不同解剖位置的信噪比和对比噪声比均无统计学显著差异。
  • 模体测量显示任一协议的有效剂量均不超过0.33 mSv。
  • 对四种低剂量协议共86次热释光剂量计测量的总体评估显示辐射剂量存在统计学显著差异(p < 0.0001),双源协议辐射剂量最低。

诊断标准与关键数值

  • 协议可诊断性由两名有经验放射科医师依据解剖结构的清晰度和锐利度判定为可诊断。
  • 图像质量评估包括噪声、空间分辨率、诊断可接受性和伪影,四种协议均评为最佳。
  • 客观图像质量通过测量不同解剖位置的信噪比和对比噪声比确定,四种协议间无统计学显著差异。
  • 有效剂量由模体热释光剂量计测量确定,任一协议均不超过0.33 mSv。

临床应用与阅片要点

  • 原文结论认为亚毫希沃特腹部CT可行且图像质量良好,剂量甚至低于常规腹部X线摄影。
  • 原文指出双源协议达到最低剂量,进一步表明双源成像可在不增加剂量的情况下进入亚毫希沃特范围。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. The purpose of this study was to evaluate radiation dose, effective dose, and image quality of different low-dose abdominal CT protocols in a swine model and an anthropomorphic phantom using a third-generation dual-source CT scanner. MATERIALS AND METHODS. Four different abdominal low-dose protocols were established using a swine model and were regarded as diagnostic by two experienced radiologists on the basis of clarity and sharpness of anatomic structures. General image conditions such as noise and spatial resolution as well as diagnostic acceptability and artifacts were evaluated. Objective image quality was determined by measuring signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in different anatomic locations. To evaluate the effective dose, thermoluminescent dosimeter (TLD) measurements were repeated in a phantom. RESULTS. Diagnostic acceptability, spatial resolution, and noise were rated as optimal in all four protocols, which were therefore regarded as diagnostic. We found no statistically significant differences in SNR or CNR for the four low-dose protocols. Effective dose determined from the phantom measurements did not exceed 0.33 mSv for any protocol. Overall evaluation of the 86 TLD measurements for the four low-dose protocols revealed a statistically significant difference in radiation dose (p < 0.0001), showing that the dual-source protocol had the lowest radiation dose. CONCLUSION. Submillisievert abdominal CT is feasible with good image quality and doses even lower than conventional abdominal radiography. Our dual-source protocol achieved the lowest dose, which further shows that dual-source imaging is possible in the submillisievert range without additional dose.