复杂混杂因素校正的R2*映射用于MRI肝脏铁定量
这项前瞻性研究纳入40例疑似铁过载患者和10名健康对照,在1.5 T和3.0 T下采用四种R2*采集方案并与FDA批准的R2法LIC参考值比较。复杂混杂因素校正R2*在采集方案间可重复性高(108次两两比较中96次P>0.05,各场强/肝段ICC>0.91),对四个临床相关LIC阈值的AUC均>0.95,并得出1.5 T与3.0 T下R2*校准LIC的公式。
要点
- 复杂混杂因素校正R2*在1.5 T和3.0 T、不同空间分辨率、回波时间和层面方向下均显示高可重复性。
- 各采集方案与基于R2的LIC定量高度相关,四个临床相关LIC阈值AUC均>0.95。
- 研究给出1.5 T(LIC=-0.04+2.62×10⁻²R2*)和3.0 T(LIC=0.00+1.41×10⁻²R2*)的R2*校准LIC公式。
核心结论
- 复杂混杂因素校正R2*映射可在1.5 T和3.0 T下以高采集方案间可重复性实现LIC定量。
研究设计与样本
- 前瞻性研究,经IRB批准并获书面知情同意,纳入40例疑似铁过载患者和10名健康对照。
- 每位受试者在1.5 T和3.0 T下各采集四种不同R2*映射方案,并以1.5 T下FDA批准的基于R2的方法作为LIC参考。
- R2*图由复数数据重建,包括噪声效应和脂肪信号校正;在每个受试者、场强和R2*采集下于九个Couinaud肝段和脾脏进行R2*测量。
关键影像表现/征象
- R2*在采集方案间显示高可重复性:2个场强和9个肝段共108次两两比较中96次P>0.05,每个场强/肝段组合ICC>0.91。
- R2*对四个临床相关LIC阈值具有高预测能力,AUC均>0.95。
- R2*与LIC的校准关系为1.5 T下LIC=-0.04+2.62×10⁻²R2*,3.0 T下LIC=0.00+1.41×10⁻²R2*。
诊断标准与关键数值
- 以1.5 T下FDA批准的基于R2的方法测得的LIC作为参考标准。
- 复杂混杂因素校正R2*对四个临床相关LIC阈值的AUC均>0.95。
- 校准公式:1.5 T下LIC=-0.04+2.62×10⁻²R2*,3.0 T下LIC=0.00+1.41×10⁻²R2*,用于将R2*换算为LIC。
临床应用与阅片要点
- 原文结论认为复杂混杂因素校正R2*映射可在1.5 T和3.0 T下以高采集方案间可重复性实现LIC定量。
- 原文指出混杂因素校正R2*在包括不同空间分辨率、回波时间和层面方向的采集方案下均可提供可重复的R2*。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: MRI-based R2* mapping may enable reliable and rapid quantification of liver iron concentration (LIC). However, the performance and reproducibility of R2* across acquisition protocols remain unknown. Therefore, the objective of this work was to evaluate the performance and reproducibility of complex confounder-corrected R2* across acquisition protocols, at both 1.5 T and 3.0 T.
METHODS: In this prospective study, 40 patients with suspected iron overload and 10 healthy controls were recruited with IRB approval and informed written consent and imaged at both 1.5 T and 3.0 T. For each subject, acquisitions included four different R2* mapping protocols at each field strength, and an FDA-approved R2-based method performed at 1.5 T as a reference for LIC. R2* maps were reconstructed from the complex data acquisitions including correction for noise effects and fat signal. For each subject, field strength, and R2* acquisition, R2* measurements were performed in each of the nine liver Couinaud segments and the spleen. R2* measurements were compared across protocols and field strength (1.5 T and 3.0 T), and R2* was calibrated to LIC for each acquisition and field strength.
RESULTS: R2* demonstrated high reproducibility across acquisition protocols (p > 0.05 for 96/108 pairwise comparisons across 2 field strengths and 9 liver segments, ICC > 0.91 for each field strength/segment combination) and high predictive ability (AUC > 0.95 for four clinically relevant LIC thresholds). Calibration of R2* to LIC was LIC = - 0.04 + 2.62 × 10-2 R2* at 1.5 T and LIC = 0.00 + 1.41 × 10-2 R2* at 3.0 T.
CONCLUSIONS: Complex confounder-corrected R2* mapping enables LIC quantification with high reproducibility across acquisition protocols, at both 1.5 T and 3.0 T.
KEY POINTS: • Confounder-corrected R2* of the liver provides reproducible R2* across acquisition protocols, including different spatial resolutions, echo times, and slice orientations, at both 1.5 T and 3.0 T. • For all acquisition protocols, high correlation with R2-based liver iron concentration (LIC) quantification was observed. • The calibration between confounder-corrected R2* and LIC, at both 1.5 T and 3.0 T, is determined in this study.