使用kat-ARC加速序列对法洛四联症修复术后患者右心室容积与功能的MRI评估

MRI Assessment of Right Ventricular Volumes and Function in Patients With Repaired Tetralogy of Fallot Using kat-ARC Accelerated Sequences.
PubMed 文献 · AJR Am J Roentgenol · 2020-08-05

前瞻性纳入2017年3月至2018年7月法洛四联症修复术后行心脏MRI的17例患者(最终分析15例),以2D cine bSSFP为参考标准,比较3D cine kat-ARC与4D flow kat-ARC序列测量心室容积与功能。3D cine kat-ARC序列测量右心室容积与功能准确性和可重复性好;4D flow序列高估容积导致EF低估,但其短扫描时间有应用前景。

要点

  • 3D cine kat-ARC与参考标准2D bSSFP在右心室EDV、ESV相关性好(r=0.94-0.98),EF相关性r=0.79。
  • 4D flow kat-ARC高估心室容积并低估EF(EDV偏差-10.2 mL,ESV偏差-22.7 mL,EF偏差6.4%)。
  • 观察者间一致性:心室容积ICC 0.93-0.99,EF的ICC 0.59-0.83。

核心结论

  • 与参考标准2D bSSFP序列相比,3D cine kat-ARC加速序列在法洛四联症修复术后患者右心室测量中显示出良好的准确性和可重复性;4D flow kat-ARC序列扫描时间短,用于容积测量似乎有前景。

研究设计与样本

  • 前瞻性研究,初始研究人群为2017年3月至2018年7月行心脏MRI(含钆对比剂注射)的法洛四联症修复术后随访患者17例,最终研究队列为其中15例;在3D cine kat-ARC、4D flow kat-ARC和2D cine bSSFP序列上测量心室容积,以2D cine bSSFP序列为心脏评估参考标准,行技术间和观察者间分析,采用Bland-Altman分析和相关性检验比较定量测量。

诊断标准与关键数值

  • 右心室EDV和ESV在3D与4D序列均相关性好(r=0.94-0.98)。
  • EF在3D序列r=0.79、4D序列r=0.92。
  • Bland-Altman分析显示左右心室容积一致性良好,左心室一致性界限更窄,组内相关系数(ICC)大于0.80。
  • 4D flow序列心室容积被高估,导致EF被低估(EDV偏差=-10.2 mL,ESV偏差=-22.7 mL,EF偏差=6.4%)。
  • 观察者间一致性:心室容积ICC 0.93-0.99,EF的ICC 0.59-0.83。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE. The objective of our study was to assess the accuracy and reproducibility of right ventricular volumes and function measurements in patients with repaired tetralogy of Fallot using two k-adaptive-t autocalibrating reconstruction for cartesian sampling (kat-ARC) accelerated sequences: a morphologic 3D cine sequence and a functional free-breathing 4D flow sequence. SUBJECTS AND METHODS. Seventeen patients who underwent cardiac MRI with gadolinium injection as part of follow-up of repaired tetralogy of Fallot from March 2017 to July 2018 were prospectively included in the initial study population; the final study cohort was composed of 15 of the 17 patients. Ventricle volume measurements were performed on a 3D cine kat-ARC sequence, a 4D flow kat-ARC sequence, and a 2D cine balanced steady-state free precession (bSSFP) sequence. The 2D cine bSSFP sequence was the reference standard in cardiac assessment. Intertechnique and interobserver analyses were performed. Bland-Altman analysis and correlation tests were used to compare quantitative measurements. RESULTS. Ventricular end-diastolic volume (EDV) and end-systolic volume (ESV) values were well correlated in the right ventricle (r = 0.94-0.98) for both 3D and 4D sequences. Ejection fraction (EF) also showed good correlation for both 3D and 4D sequences (r = 0.79 and r = 0.92). Bland-Altman analysis showed good agreement between right and left ventricular volumes, with narrower limits of agreement in the left ventricle, and an intraclass correlation coefficient (ICC) of greater than 0.80. For the 4D flow sequence, ventricular volumes were overestimated, which led to underestimation of the EF (bias for EDV = -10.2 mL, bias for ESV = -22.7 mL, bias for EF = 6.4%). Interobserver agreement was excellent for the ventricular volumes (ICC, 0.93-0.99) and fair to excellent for the EFs (ICC, 0.59-0.83). CONCLUSION. Compared with the reference standard 2D bSSFP sequence, the 3D cine kat-ARC accelerated sequence showed good accuracy and reproducibility for right ventricular measurements in patients with repaired tetralogy of Fallot. The short duration of the 4D flow kat-ARC sequence appears promising for performing volumetric measurements.