概念验证:体模研究确保COVID-19疫情期间隔玻璃行便携式胸部X线摄影的质量与安全

Proof of Concept: Phantom Study to Ensure Quality and Safety of Portable Chest Radiography Through Glass During the COVID-19 Pandemic.
PubMed 文献 · Invest Radiol · 2021-01-01

本研究使用成人拟人胸部体模,在无玻璃、智能玻璃和普通玻璃条件下采集16幅胸片,由2名放射科医生按5分Likert量表独立评分图像质量(量子斑驳和结构锐利度)并测量入射表面剂量。结果显示智能玻璃在100 kVp 12 mAs、20 mAs及125 kVp 6.3 mAs下图像质量评分等于或高于常规摄影,图像质量评分与入射表面剂量在各条件间无统计学显著差异,智能玻璃后向散射在3 m处极小、4 m处为零。结论为通过隔离室智能玻璃门可安全获得良好质量的便携式胸片,但存在轻微后向散射辐射,可能需要调整曝光参数(如增加毫安秒)以优化图像质量。

要点

  • 智能玻璃在100 kVp 12 mAs、20 mAs及125 kVp 6.3 mAs下图像质量评分等于或高于常规摄影。
  • 无玻璃、智能玻璃和普通玻璃三组间图像质量评分与入射表面剂量均无统计学显著差异。
  • 智能玻璃后向散射在距X线管3 m处极小,4 m处记录为零。

核心结论

  • 通过隔离室智能玻璃门可安全获得良好质量的便携式胸部X线摄影图像,但该技术会产生轻微后向散射辐射。

研究设计与样本

  • 体模研究,使用成人拟人胸部体模,共采集16幅胸片,采用不同曝光条件在无玻璃、智能玻璃和普通玻璃下成像,由2名放射科医生独立评分。

诊断标准与关键数值

  • 图像质量采用5点Likert量表评分,评估量子斑驳和解剖结构锐利度。
  • 智能玻璃条件下图像质量评分等于或高于常规摄影,所用曝光条件为100 kVp 12 mAs、100 kVp 20 mAs和125 kVp 6.3 mAs;常规摄影条件为100 kVp 2 mAs和125 kVp 3.2 mAs。
  • 图像质量Likert评分和入射表面剂量在无玻璃、智能玻璃和普通玻璃三组间均无统计学显著差异。

临床应用与阅片要点

  • 作者指出使用该技术时可能需要调整曝光参数(如增加毫安秒)以优化图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Chest radiography is often used to detect lung involvement in patients with suspected pneumonia. Chest radiography through glass walls of an isolation room is a technique that could be immensely useful in the current COVID-19 pandemic.

PURPOSE: The purpose of this study was to ensure quality and radiation safety while acquiring portable chest radiographs through the glass doors of isolation rooms using an adult anthropomorphic thorax phantom.

MATERIALS AND METHODS: Sixteen chest radiographs were acquired utilizing different exposure factors without glass, through the smart glass, and through regular glass. Images were scored independently by 2 radiologists for quantum mottle and sharpness of anatomical structures using a 5-point Likert scale. Statistically significant differences in Likert scale scores and entrance surface dose (ESD) between images acquired without glass and through the smart and regular glass were tested. Interreader reliability was also evaluated.

RESULTS: Compared with conventional radiography, equal or higher mean image quality scores (mottle and anatomical structures) were observed with the smart glass using 100 kVp at 12 mAs and 20 mAs and 125 kVp at 6.3 mAs (100 kVp at 2 mAs and 125 kVp at 3.2 mAs were used for conventional radiography observations). There was no statistically significant difference in the Likert scale scores for image quality and the entrance surface dose for radiographs acquired without glass, through the smart glass, and through regular glass. Backscatter from the smart glass was minimal at a distance of 3 m and was recorded as zero at a distance of 4 m from the x-ray tube outside an isolation room.

CONCLUSIONS: Good-quality portable chest radiographs can be obtained safely through the smart glass doors of the isolation room. However, this technique does result in minor backscatter radiation. Modifications in the exposure factors (such as increasing milliampere seconds) may be required to optimize image quality while using this technique.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Rai A, MacGregor K, Hunt B, et al. Proof of Concept: Phantom Study to Ensure Quality and Safety of Portable Chest Radiography Through Glass During the COVID-19 Pandemic. Invest Radiol. 2021. PMID: 32773486. doi:10.1097/RLI.0000000000000716

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。