不同病灶与CT成像条件下影像组学特征最小可检测差异的量化
本研究通过模拟297种成像条件下的人体模肺结节模型图像,计算19个形态学特征的最小可检测差异,发现其范围为1%至175%,并显示QIBA协议推荐可改善最小可检测差异,进一步将层厚限制在0.5 mm至1 mm可能进一步改善。
要点
- 影像组学特征的最小可检测差异随特征和成像协议不同,范围为1%至175%。
- QIBA协议推荐相比可能的协议范围可改善最小可检测差异。
- 将层厚限制在0.5 mm至1 mm可能进一步改善最小可检测差异。
诊断标准与关键数值
- 最小可检测差异范围为1%至175%,取决于具体特征和成像协议组合。
- QIBA协议推荐相比可能的协议范围可改善最小可检测差异。
- 将层厚要求进一步限制在0.5 mm至1 mm可能改善最小可检测差异。
临床应用与阅片要点
- 研究结果可前瞻性用于指导成像协议决策,以提供更优的影像组学特征量化。
- 作者指出QIBA协议推荐可改善最小可检测差异。
- 作者指出将层厚限制在0.5 mm至1 mm可能进一步改善最小可检测差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: The 3-fold purpose of this study was to (1) develop a method to relate measured differences in radiomics features in different computed tomography (CT) scans to one another and to true feature differences; (2) quantify minimum detectable change in radiomics features based on measured radiomics features from pairs of synthesized CT images acquired under variable CT scan settings, and (3) ascertain and inform the recommendations of the Quantitative Imaging Biomarkers Alliance (QIBA) for nodule volumetry.
MATERIALS AND METHODS: Images of anthropomorphic lung nodule models were simulated using resolution and noise properties for 297 unique imaging conditions. Nineteen morphology features were calculated from both the segmentation masks derived from the imaged nodules and from ground truth nodules. Analysis was performed to calculate minimum detectable difference of radiomics features as a function of imaging protocols in comparison to QIBA guidelines.
RESULTS: The minimum detectable differences ranged from 1% to 175% depending on the specific feature and set of imaging protocols. The results showed that QIBA protocol recommendations result in improved minimum detectable difference as compared to the range of possible protocols. The results showed that the minimum detectable differences may be improved from QIBA's current recommendation by further restricting the slice thickness requirement to be between 0.5 mm and 1 mm.
CONCLUSION: Minimum detectable differences of radiomics features were quantified for lung nodules across a wide range of possible protocols. The results can be used prospectively to inform decision-making about imaging protocols to provide superior quantification of radiomics features.
引用
- Hoye J, Solomon JB, Sauer TJ, et al. Quantification of Minimum Detectable Difference in Radiomics Features Across Lesions and CT Imaging Conditions. Acad Radiol. 2020. PMID: 32828664. doi:10.1016/j.acra.2020.07.029
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。